BEFORE PLUGGING THIS INSTRUMENT into the Mains (line) voltage,
be sure the correct voltage and fuse have been selected.
Verify that the line voltage selection card and the fuse are matched to the power source. Refer to
Figure 2-1, Line Voltage and h e Selection.
h s e s may be ordered under HP part numbers 2110-0083, 2.5A (250V normal blow) for 115 Vac
operation and 2110-0043, 1.5A (250V normal blow) for 230 Vac operation.
Operating voltage is shown
in module window.
Selection of Operating Voltage and Fuse Replacement
1.
Open cover door,
pull
the
FUSE
PULL
lever and rotate to left. Remove the fuse.
2.
Remove the Line Voltage Selection Card. Position the card so the line voltage appears at top-left
3.
Rotate the
FUSE
PULL
lever to
its
normal position. Insert a fuse
of
the correct value in the holder.
corner.
Push
the card
firmly
into the slot.
Close the cover door.