HP 8901B Operation And Calibration Manual page 197

Hide thumbs Also See for 8901B:
Table of Contents

Advertisement

Operation
Model 8901B
4.
Enter the new frequency/cal factor pairs by selecting
I
E n t e r To b
1
e
1.
a. The controller will prompt:
"New Cal Factor
f o r
R E F position c a l
f a c t o r ? "
Enter the reference calibration factor. For example, enter 100% as:
1
0 0 <ENTER>
b. The controller will prompt:
"New Frequency f o r number '
1
'
cal factor?"
Enter the frequency (in MHz) for the first frequency/cal factor pair. For example, enter
100 kHz as:
.
1
<ENTER>
c. The controller will prompt:
"New Cal
Factor for
number
'1'
frequency?"
Enter the cal factor for the first frequency/cal factor pair, for example, enter 99.0% as:
9 9 . 0 <ENTER>
d. Continue entering the nine other frequency/cal factor pairs until the table is complete.
Note that selecting
ENTER
as a response to a prompt withour specifying a value
will
automatically
select the value currently in the table.
Select
D
i
I
P
1
ay
Tab 1
e
I
to see the cal factors that you have entered.
Select
I
Send
Analyzer.
I
to put those cal factors into the cal factor table in the Modulation
To display the cal factors currently used by the Modulation Analyzer, select
I
Read Cal
I .
The controller will query the Modulation Analyzer and then display a list of frequency/cal-factor
pairs currently in the table. The cal factors are listed in the format
50
. 1
-3
1
.o
3.0
10
30
100
300
1000
2600
98.9
99.0
97.2
99.8
99.5
98.7
99.3
97.9
98.1
99.1
99.7
The first column lists each frequency (in MHz) of the frequency/cal factor pair. The second
column lists each cal factor (in
%)
of the frequency/cal factor pair. (The first frequencyhal factor
pair listed is the reference cal factor that
is
measured a t 50 MHz. This frequency cannot be
changed.) To enter new frequency/cal factor pairs,
go to
step
4.
3-82
RF Power Calibration Factor Entry Example Program