Replace the Solvent Mixer

Replace the Solvent Mixer

Parts required
p/n
G1312-87330
1
Lift up solvent filters in solvent reservoirs to avoid
leakages.
2
Open the purge valve.
3
Power off the system.
4
Open the doors.
6
Remove the solvent mixer.
8
Install the capillaries to the solvent mixer.
Preparative Binary Pumps User Manual
Description
Mixer
5
Remove the capillaries from the solvent mixer.
7
Install the solvent mixer.
Next Steps:
9
Make sure all capillary and tubing connections are
reconnected and tight.
10 Insert solvent filters into solvent reservoirs.
11 Power on the system.
12 Purge the system.
13 Close the doors.