SERVICE MANUAL
Chemical Replenishment
[1] Check the replenishment rates. Use the procedure on Page 31.
[2] Check the STRAINERS for debris:
(a) Place CLAMPS on the TUBING to stop the flow of replenishment solution.
(b) Disassemble the STRAINER AYs.
(c) Use a BRUSH and warm water to clean the STRAINER AYs and the SCREENS.
INTERLOCK SWITCH
MAGNET
[1] Press the RUN/STANDBY SWITCH.
Warning
Moving Parts
[2] Remove the TOP COVER.
[3] Check that the components stop operating:
• MAIN DRIVE MOTOR
• BLOWER MOTOR
• DRYER HEATER
[4] If the components do not stop when the TOP COVER is removed, install a new INTERLOCK SWITCH.
52
STRAINER ASSEMBLY
TOP COVER
MAGNET
INTERLOCK
SWITCH
SCREEN
H172_1063BCA
H172_1063BA
H172_1064BCA
H172_1064BC
12JAN99 – 7C8772