Kodak X-Omat 2000 Service Manual page 39

Hide thumbs Also See for X-Omat 2000:
Table of Contents

Advertisement

2 GASKETS
WATER INLET
TUBE
[11] Move the RACK SUPPORT toward the non-drive side of the PROCESSOR.
[12] Install the 2 SCREWS.
[13] Fill the TANKS with water.
[14] Check for leakage.
[15] Drain the water from the DEVELOPER TANK and the FIXER TANK.
[16] Fill the DEVELOPER TANK and the FIXER TANK with the correct solution.
[17] Connect the water supply.
Warning
Dangerous Voltage
[18] Connect the main power.
7C8772 – 12JAN99
2 FITTINGS
TANK
HEATER
EXCHANGER
2 O-RINGS
2 WASHERS
2 GLAND NUTS
H172_1054CCA
H172_1054CA
Adjustments and Replacements
[7] Remove if necessary:
• 2 FITTINGS
• WATER INLET TUBE
[8] To remove the TANK, lift the non-drive side of
the TANK first.
[9] Remove from the TANK:
• 2 GLAND NUTS
• 2 O-RINGS
• 2 WASHERS
• HEAT EXCHANGER
Important
Use new GASKETS if you are installing a FIXER
TANK.
[10] Install:
• new HEAT EXCHANGER
• 2 O-RINGS
• 2 WASHERS
• 2 GLAND NUTS
• TANK
• 2 FITTINGS
• 2 GASKETS
• WATER INLET TUBE
• TANK SEALS
• TANK CLIP
• TUBING
• RACK SUPPORT
39