Shipping problems?

If the shipping materials are damaged or the contents of the container are incomplete:
• Contact the nearest Agilent Technologies office.
• Keep the shipping materials for the carrier's inspection.
• If you must return an test set to Agilent Technologies, use the original (or

Test Set Location and Rack Mounting Requirements

Locating the test set
Make sure that the fan inlet and exhaust vent areas on the sides of the test set are not
obstructed. The minimal required clearance is 2 inches. Airflow restrictions cause
additional airflow noise and cause the fans to speed up so they can draw in enough air for
the required cooling resulting in excessive audible noise.
Cooling and rack mounting
Do not rack mount the test set side-by-side with any other instrument with side-by-side
ventilation. Make sure the exhaust air from the first instrument is directed away from the
inlet of the second unit. If the pre-heated air from the first instrument is directed into the
second instrument, it can cause excessive operating temperatures in the second unit and
can cause instrument failures. The test set draws air in from the left side and exhausts air
from the right side.
14
comparable) shipping materials. See
"Returning a test set for Service"
on page 101.