1
Quick-Start
Steps
8 Verify the installation
9 Attach a loopback cable
10 Generate and view a
signal
18
Actions
a From the mouse right-click
menu, select
,
Utility
Virtual
Front Panel
. (All
key-presses that follow
refer to the keys shown on
the Virtual Front Panel.)
b Press
,
,
System
Show
System
.
c Verify that the purchased
application(s) appear in the
list.
a Use an RF cable to connect
the front panel RF In and
RF Out ports for this
sub-instrument.
a Press
,
Input/Output
More
RF Output & Test Set
Config
,
RF Output,
.
RF Output
b Press
AMPTD Y Scale
,
,
,
.
Value
10
dBm
c Press
,
FREQ
Center Freq
2.4
,
GHz
.
d Press
,
SPAN X Scale
Span
,
8
,
MHz
.
e Press
Source, Frequency,
2.4
,
GHz
.
f Press
,
Source
RF Output
.
On
g Press
,
Source
Amplitude
,
,
.
RF Power
0
dBm
Notes
If you require further assistance, contact the Agilent
support team.
Online assistance:
The purpose is to use a loopback cable to verify that
the test set's source can generate a signal and the test
set's analyzer can measure it.
,
Select RF Output rather than GPS Output.
Set the amplitude reference to 10 dBm
Set the analyzer center frequency to 2.4 GHz.
Set the analyzer span to 8 MHz.
Ref
Set the source frequency to 2.4 GHz.
Set the source RF output to On.
,
Set the source amplitude to 0 dBm.
To ensure accurate power levels from the source, see
"Source Level Accuracy"
,
,
http://www.agilent.com/find/assist
on page 20.
Getting Started Guide