Delay Volume And Extra-Column Volume; Delay Volume - Agilent Technologies 1290 Infinity LC System System Manual And Quick Reference

Table of Contents

Advertisement

3
Optimization of the Agilent 1290 Infinity LC System

Delay Volume and Extra-Column Volume

Delay Volume and Extra-Column Volume

The delay volume is defined as the system volume between the point of mixing
in the pump and the top of the column.
The extra-column volume is defined as the volume between the injection point
and the detection point, excluding the volume in the column.

Delay Volume

In gradient separations, this volume causes a delay between the mixture
changing in the pump and that change reaching the column. The delay
depends on the flow rate and the delay volume of the system. In effect, this
means that in every HPLC system there is an additional isocratic segment in
the gradient profile at the start of every run. Usually the gradient profile is
reported in terms of the mixture settings at the pump and the delay volume is
not quoted even though this will have an effect on the chromatography. This
effect becomes more significant at low flow rates and small column volumes
and can have a large impact on the transferability of gradient methods. It is
important, therefore, for fast gradient separations to have small delay
volumes, especially with narrow bore columns (e.g., 2.1 mm i.d.) as often used
with mass spectrometric detection.
The delay volume in a system includes the volume in the pump from the point
of mixing, connections between pump and autosampler, volume of the flow
path through the autosampler and connections between autosampler and
column.
As an example, in HPLC methods using 5 µm packing material flow rates of
1 ml/min are typically used in a 4.6 mm i.d. column and about 0.2 ml/min in a
2.1 mm i.d column (same linear velocity in the column). On a system with a
typical delay volume of 1000 µl and using a 2.1 mm column there would be an
initial "hidden" isocratic segment of 5 min whereas on a system with 600 µl
delay volume the delay would be 3 min. These delay volumes would be too high
for run times of one or two minutes. With sub-two µm packings the optimum
flow rate (from the Van Deemter Curve) is a little higher and so fast
chromatography can use three to five times these flow rates yielding delay
40
Agilent 1290 Infinity LC System Manual and Quick Reference

Advertisement

Table of Contents
loading

Table of Contents