Step 5 – Summary
Check parameter settings, and click BACK to return and revise values, or TRANSFER to
transfer values to the device.
The message Quick Setup was successful will appear. Click OK.

Linearization

You can use the linearization feature to define a more complex vessel shape and enter up
to 11 level breakpoints where the corresponding volume is known. The values
corresponding to 100% and 0% levels must be entered. The breakpoints can be ordered
from top to bottom, or the reverse.
Example (values are for example purposes only.)
Breakpoint
Level
number
value
4
3
2
1
7ML19985JR01
SITRANS LR200 (PROFIBUS PA) – INSTRUCTION MANUAL
Level
Break-
value
point no.
(m)
19.5
1
0
2
5
9
3
9
5
4
19.5
0
0.5 m
Volume
value
(l)
0
20 m
500
3000
8000
Page 33