Linearization; Using Linearization Via The Quick Start Wizard - Siemens LR250 Instruction Manual

Sitran sradar level transmitter
Hide thumbs Also See for LR250:
Table of Contents

Advertisement

Step 3 – Vessel Shape
1)
Click on Step 3.
2)
Select a predefined vessel shape. (To describe a more complex shape see

Linearization

3)
Click on Apply.
Step 4 – Ranges
1)
Click on Step 4.
2)
Change units if desired (in meters by default).
3)
Set High and Low Calibration points.
4)
To convert the reading from level to volume enter a value for Maximum Volume.
5)
If a vessel shape with parabolic ends has been selected, set dimensions A and L.
6)
Click on Apply.
Step 5 – Summary
Check parameter settings, and click on Cancel to abort, or Apply to transfer values to the
device.
Linearization
You can use the linearization feature to define a more complex vessel shape and enter up to
32 level breakpoints where the corresponding volume is known. The values corresponding to
100% and 0% levels must be entered. The breakpoints can be ordered from top to bottom, or
the reverse.

Using Linearization via the Quick Start wizard

1)
Navigate to Configure/Setup > Wizard > Quick Start.
2)
In Step 1 – Identification, make any desired modifications and click on Apply.
3)
In Step 2 – Application, select a level application and liquid as a material.
4)
In Step 3 – Vessel Shape, choose the vessel shape option Linearization Table.
5)
In Step 4 – Ranges, enter a value for maximum volume.
7ML19985JE06
below.)
max. volume
SITRANS LR250 (HART) – INSTRUCTION MANUAL
Breakpoints
Page 65