Inspecting And Cleaning The Cuvette Splash Covers - Siemens ADVIA 2400 Operator's Manual

Chemistry system
Hide thumbs Also See for ADVIA 2400:
Table of Contents

Advertisement

Inspecting and cleaning the cuvette splash covers

Materials required:
• Lint-free towel
• Deionized water
Time: 5 minutes
Analyzer mode: READY
NOTE:
entering the dilution and reaction cuvettes.
1. Inspect the cuvette covers for spills and splattering.
If there is any splattering is on the cuvette covers (1), proceed to step 2.
Figure 5-7. Cuvette covers
2. Using lint-free towels moistened with deionized water, wipe down the covers.
WARNING
Do not touch probes or mixing rods, to avoid contamination.
3. If splattering is extensive or enters the cuvettes, call your local technical support
provider or distributor.
Maintenance
Cuvette covers are installed around the probes to prevent water and reagent from
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.