Operating condition
Diluent vol.
Indicates the status of the liquid volume of
diluent (saline solution). Insufficient liquid
is detected by a liquid level sensor in the
bottle.
Cuv. Wash Vol.
Status of the liquid volume of cuvette wash
solution. Insufficient liquid is detected by a
liquid level sensor in the bottle.
Cuv. Cond. Vol.
Status of the liquid volume of cuvette
conditioner. Insufficient liquid is detected
by a liquid level sensor in the bottle.
Conc.wast.tank (if installed)
Status of the volume of the optional
concentrated waste tank. A sensor monitors
whether the tank is filled with waste.
Waste tank
Status of the volume of the optional waste
tank. A sensor monitors whether the tank is
filled with waste.
3. Checking the availability of the reagents and wash solutions.
Visually check the system reagents.
a. Perform a prime after replacing any system ancillary reagents.
b. At the Operation Panel, select
c. Select
d. Visually check the ISE reagents.
e. Perform a prime after replacing the ISE Buffer:
(1) At the Menu Panel, select
(2) In the Times box of the Bufferprime area, type the number of prime cycles
(3) Select
f.
Visually check the controls and calibrators on the calibrator/control tray.
g. Visually check the wash solutions.
h. Visually check the lamp coolant level.
Operating the System
, then select
PRIME 2
Execute
you want (
is recommended).
20
.
Execute
Normal
Abnormal indicator and corrective
indicator
action
OK
NG
Fill the diluent bottle with 0.9% saline
solution.
Make sure the sensor is in the proper
position.
OK
NG
Fill the cuvette wash solution bottle.
Make sure the sensor is in the proper
position.
OK
NG
Fill the cuvette conditioner bottle. Make
sure the sensor is in the proper position.
OK
NG
Empty the concentrated waste tank.
If there is still a problem, check the tubes
for obstructions.
OK
NG
Verify that the waste tubes are free of
obstructions. If the problem persists,
contact your local support provider or
distributor.
.
Prime
.
., then select
Maint
ISE Operation
.
69