6. Fasten the connector.
7. Using a pad soaked in alcohol, clean the outside surfaces of the filter holders and
hoses.
8. Insert the filter hoses into the bottles, then fasten the caps.
9. Return the analyzer to the READY mode.
10. At the Operation Panel, select Prime.
11. In the PRIME Set dialog box, select Prime 2 and enter 10 or more for the number of
times in all fields, and then select Execute.
Cleaning the pure-water bottle filter
Materials required:
• Ten 10R filters (REF 01160530,
PN 073-0035-01)
• One 18R filter (REF 01448895,
PN 073-0034-01)
• Hex wrench
Time: 15 minutes
Analyzer mode: READY
Clogged filters create an insufficient flow rate and produce air bubbles.
NOTE:
the filter with the one in the kit, resume operation, and then clean and store the removed
filter for the next scheduled maintenance.
Figure 5-24. Pure-water bottle
1. Ensure that the instrument is in READY mode.
2. Remove the silicon return hose (1) from the top front of the pure water bottle.
The filter is contained in a metal filter holder attached to the end of the return hose.
138
A set of filters is included in the supplies kit. To avoid system down-time, replace
BIOHAZARD
Wear personal protective
equipment.
Use universal precautions.
1)
Pure-water bottle
ADVIA 2400 Operator's Guide