Washing The Lines; Rinsing The Lines - Siemens ADVIA 2400 Operator's Manual

Chemistry system
Hide thumbs Also See for ADVIA 2400:
Table of Contents

Advertisement

Figure 5-37. ISE electrode plate thumbscrew
7. Remove the electrodes and replace it with the dummy electrode.
8. At the ISE Operation window, in the Bufferprime area, type
9. Select
10. When prompted, select

Washing the lines

1. Remove the buffer bottle with the deionized water and replace it with a bottle filled
with a solution of 475 mL of deionized water and 25 mL probe wash 3 solution.
2. At the ISE Operation window, in the Bufferprime area, enter
3. Select
4. When prompted, select

Rinsing the lines

1. Replace the probe wash 3 solution bottle with a bottle of deionized water.
2. In the Bufferprime area, enter
3. Remove the dummy electrode.
4. Reinstall the Na, K, and Cl electrodes.
5. In the Initialize area, select
6. Before reinstalling the buffer-solution bottle, thoroughly rinse the buffer bottle cap,
float switch, and tube with deionized water and dry completely.
7. Install the buffer bottle or replace it if the volume is low.
156
.
Execute
to execute buffer prime.
Yes
.
Execute
to execute the buffer prime.
Yes
Execute
in the Times field, then select
50
.
in the Times field.
50
in the Times field.
50
.
Execute
ADVIA 2400 Operator's Guide