Replacing Dpp Probes Equipped With Crash Detection; Removing The Dpp Probe - Siemens ADVIA 2400 Operator's Manual

Chemistry system
Hide thumbs Also See for ADVIA 2400:
Table of Contents

Advertisement

10. At the Operation Panel, select
wash cups).
11. At the Operation Panel, select
proper water flow through the probe.
NOTE:

Replacing DPP probes equipped with crash detection

Materials required:
• Probes:
DPP
• Phillips screwdriver
• Pliers
• Lint-free towels
Time: 10 minutes
Analyzer mode: STANDBY
Figure 5-31. Dilution probe location

Removing the DPP probe

1. Put the system in Standby mode.
CAUTION
Manually support the probe and be careful not to strike it against anything on the
analyzer, to avoid damaging the probe tip when the power is off.
2. Cover the cuvettes, wash cups, and other analyzer surfaces with lint-free towels to
catch any screws that may fall.
3. Lift and manually rotate the probe over the sample tray, either over the sample tray or
over the ISE.
4. Loosen but do not remove the screws on each side of the probe cover.
Maintenance
Make sure that no water is leaking from the knurled fitting.
-equipped for crash detection
(REF 02030495, PN 073-0611-01)
to return the probes back to home (over the
Initialize
,
then select
PRIME
PRIME 2
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
,
then
to ensure
Execute
1)
Dilution Probe (DPP)
2)
Sample Probe (SPP)
3)
Reagent Probe 1
(RPP1)
4)
Reagent Probe 2
(RPP2)
147