Installation Of Transit Coupling - Invacare Robin Installation And Technical Manual

Hide thumbs Also See for Robin:
Table of Contents

Advertisement

5.9.2 Installation of transit coupling

The coupling is delivered assembled with motor box, fixtures for pendles
and lock fixtures. The delivered transit coupling is prepared so that the hole
motor box is placed in one of the two rooms after installation.
See drawing.
The motor box and lock fixtures, have to be placed as the transit is delivered,
to ensure functionality and safety.
Fixtures for pendles can be placed freely as long as it complies with the
construction of the system. See measure sketch.
NB! It is very important that the thickness of the wall is controlled in advance.
When installing the transit coupling it is very important that the motor box is placed in the room where it was
intended when it was ordered, because the entire system is constructed under consideration of where the motor
box is placed.
Installation with pendles are described in section 5.2.2.1.
See the measure sketch for further information, if necessary.
NB! Installation of the transit coupling is only possible using pendles, therefore the coupling can only be installed
where ceiling mounting is possible.
The best result is obtained in a concrete ceiling.
When installing in wooden ceilings, it is very important that the traverse system and coupling are installed in a way
that they bend down equally and with minimum bending.
If it is necessary to strengthen the ceiling in a wooden ceiling construction, please make sure to strengthen with a
connected construction between rail system and transit coupling.
Do not install the pendles in different rafters this way the coupling and traverse system will not bend down equally
and the coupling will not work correctly.
33