5.8 Installation of rail switches

The rail switch is used where a change in direction is required during movement of the patient. The rail switch is
manually operated.
By coupling of the rail switch with both straight rails as well as curved rails, different combinations can be obtained.
Installation dimensions:
The rail switch can only be used in combination with ceiling installation and S-profile rails. The rail switch is
designed for standard ceiling bracket.
During ordering or installation of the straight or curved single rail system, it is important to bare in mind in which
direction the mounting dowels of the rail switch are facing. Only one standard rail switch is available - refer to the
illustration.
By installation in concrete ceiling, refer to section 5.2.1 (standard brackets), and section 5.5.2 (coupling of S-profile
rails).
As described in section 5.5.2, a rebore in the rails is performed when coupling the rails. This rebore is done in the
rail switch at delivery, and the necessary tube pins are supplied. When installing the rails to the switch, remember
to tighten the pre-installed locking screws on the rail switch - refer to the illustration.
In connection with installation of a rail switch onto different ceiling types or in combinations with pendles, please
contact Invacare® for further information.
After installation, the functions of the rail switch and the fastening of the end stops must be
inspected.