Resetting The Service Light; Replacing The Electric Actuator - Invacare Roze User Manual

Stand assist patient lift
Hide thumbs Also See for Roze:
Table of Contents

Advertisement

A soft cloth, dampened with water and a small amount of
mild detergent, is all that is needed to clean the patient lift.
The lift can be cleaned with non-abrasive cleaners.
Motors, control unit and mounting parts can be damaged if
the lift is cleaned any other way than stated above.
The lift must be wiped with a moistened, firmly wrung cloth
with ordinary household disinfectants. Only use disinfection
detergents approved by the facility and follow the facility
policy. For more information about the residence time
and concentration of disinfectants, please contact your
disinfectant dealer or the manufacturer of the disinfectant.

8.4 Resetting the Service Light

This procedure should only be performed by a
qualified technician.
This procedure should only be performed after
performing the steps in the Checking the Service
Light section.
1195052-B
You will hear a sound when the service light has
been reset.
The service light is now reset to indicate a service
requirement after a 12 month period or 8000 cycles.
If the service light is required to be set for a period
shorter than 12 months, a special hand control and
further instructions are required. Contact Invacare to
order this hand control and instructions.

8.5 Replacing the Electric Actuator

This procedure should only be performed by a
qualified technician. If possible, use an assistant
when performing this procedure.
Maintenance
1. Locate the hand control
A.
2. Press and hold the UP
button B and DOWN
button C at the same
time for five seconds.
51