Invacare® Roze™
SYMPTOMS
Unusual noise from lift arm actuator.
Lift arms will not lower in uppermost
position.
Legs don't open and close properly.
Service light is flashing
If problems are not remedied by the suggested means, please contact your dealer or Invacare.
44
FAULTS
Actuator is worn or damaged or spindle
is bent.
Lift arms require a minimum weight load
to lower from the uppermost position.
Leg actuators may be worn or damaged
or spindle is bent.
12 months have passed since the lift was
manufactured or initially assembled.
SOLUTION
Refer to Replacing the Electric Actuator
or Replacing the Leg Actuators. Contact
your Dealer or Invacare representative.
Pull down slightly on the lift arms.
Refer to Replacing the Leg Actuators.
Initial assembly — Reset the service
light. Refer to Resetting the Service
Light.
After initial assembly — Service the lift.
Contact your local Invacare dealer.