The mast assembly A may be removed from the
base B for storage or transporting. The mast
assembly MUST be properly secured to the base
assembly before use.
1. Put the base B on a level surface so all casters C make
contact with the floor.
2. Lock the rear casters. Refer to Locking/Unlocking the
Rear Casters in the Usage section of the manual.
3. Unscrew the top knob D and remove it from the base.
4. Loosen the bottom knob E, but leave it screwed into
the base.
1195052-B
5. Connect the mast assembly cable F to the base cable
G.
WARNING!
Risk of injury
– DO NOT place hands between the mast
and base during assembly.
CAUTION!
Risk of Damage.
– DO NOT crush the connected cables
between the mast and base during
assembly.
6. Put the tube of the mast assembly H into the hole in
the base I.
7. While supporting the mast assembly, tighten the bottom
plastic knob.
8. Screw the top plastic knob into the hole in the base.
9. Reverse STEPS 3–8 to disassemble the mast from the
base.