AGFA CR 10-X User Manual page 82

Hide thumbs Also See for CR 10-X:
Table of Contents

Advertisement

82 | CR 10-X, CR 12-X, CR 15-X | Remarks for HF-emission and immunity
Conducted
high frequency
disturbance
variables in
accordance
with
IEC 61000-4-6
Radiated high
frequency
disturbance
variables in
accordance
with
IEC 61000-4-3
The field strength of stationary transmitters, such as base stations of radio
telephones, mobile broadcasts for rural areas, amateur stations, and AM and
FM radio transmitters, cannot be precisely predetermined theoretically. An
2491E EN 20150409 1543
3 V
3 V
eff
eff
150 kHz to
80 MHz
3 V/m
3 V/m
80 MHz to
2.5 GHz
Recommended protective
distance:
d = 1.2
d = 1.2
80 MHz to 800
MHz
d = 2.3
800 MHz to 2.5
GHz
With P as the rated power of
the transmitter in watts (W) in
accordance with the
manufacturer information on
the transmitter and d as the
recommended protective
distance in metres (m).
The field strength of stationary
radio transmitters is lower than
the level of the agreement at all
frequencies in accordance with
an on-site investigation.
Disruptions are possible near
devices that carry the following
symbol:

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr 12-xCr 15-x

Table of Contents