9. Place back the cassette unit.
Note that the protruding elements of the cassette unit should be
positioned correctly in line with the digitizer: if the cassette unit is
positioned too high, the protruding elements on the cassette unit may
become damaged.
10. Switch on the digitizer.
2491E EN 20150409 1543
2. Open the back of the device by turning the four fixation rings by 90
degrees:
3. Gently remove the image plate from the back of the digitizer.
4. Slide the image plate completely into the cassette.
Verify that the white phosphor side is oriented to the tube side of
the cassette and that the shutter does not scratch the image plate.
5. Close the back of the device.
6. Push the release button to detach the cassette from the cassette
unit.
Note:
Never use force to clear the jammed image plate. If it
is not possible to gently remove the image plate, call
your local service organization.
Make sure not to bend the image plate when
removing it from the device.
After a jam, the image plate can be used again if it is
not damaged.