Charger
Voltage output
charger
Voltage supply
charger
Protection class
1
IP41 or IP67 (depending on version)
charger
1
See product label and label on each electric device for correct
protection class. The lowest IP classification decides the overall
classification of the device.
IPx4: Protected against water splashed from any direction.
IP24: Protected against objects larger than 12.5 mm and
protected against water splashed from any direction.
IP44: Protected against objects larger than 1 mm and protected
against water splashed from any direction.
IP41: Protected against objects larger than 1 mm and protected
against vertically dripping water.
IP67: Dust-tight and protected against temporary immersion
into water.

10.3 Environmental conditions

Storage and
transportation
Temperature
-30 °C to +50 °C
Relative humidity
10 % to 70 %
Atmospheric
pressure
1586745-B
24 V DC, max. 0.9 A
230 V AC, 50 Hz
Operation
+5 °C to +40 °C
20 % to 90 % at 30
°C, not condensing
700 hPa to 1060 hPa
10.4 Electromagnetic compliance (EMC)
information
Medical Electrical Equipment needs to be installed and used according
to the EMC information in this manual.
This equipment has been tested and found to comply with EMC limits
specified by IEC/EN 60601-1-2 for Class B equipment.
Portable and mobile RF communications equipment can affect the
operation of this equipment.
Other devices may experience interference from even the low levels
of electromagnetic emissions permitted by the above standard. To
determine if the emission from the lift is causing the interference, run
and stop running the lift. If the interference with the other device
operation stops, then the lift is causing the interference. In such rare
cases, interference may be reduced or corrected by the following:
Reposition, relocate, or increase the separation between the
devices.

10.4.1 Guidance and manufacturer's declaration

Basic
Test
Standard
Mains Port
EN
Conducted
55011:2007
Emissions
Radiated
EN
Emissions
55011:2007
EN 61000-
3-2:2006 +
Mains Harmonics
A1:2009 +
A2:2009
Technical data
Compliance Level /
Frequency Range
Group 1 Class B
Group 1 Class B
Class A