9.5 Environmental conditions

Storage and
Transportation
Temperature
-10°C to +50°C
Relative humidity
20% to 80% - not condensing
Atmospheric
pressure
Allow the product to reach operation temperature
before usage:
Warming up from minimum storage temperature
might take up to 24 hours.
Cooling down from maximum storage
temperature might take up to 24 hours.

9.6 Electrical system

Voltage
Uin = 230 Volt, AC, 50/60 Hz (AC =
supply
Alternating current)
Maximum
Iin max. = 1.5 A
current input
Intermittent
(periodic
Int = 10%, max. 2 min / 18 min
motor
operation)
24
Operation
+5°C to +40°C
700 hPa to 1060 hPa
Degree of
IPX6 or IPX4 (depending on configuration)
protection
When the bed is equipped with electrical
leg rest it is protected according to IPX4.
IPX6 – The electrical system is protected
against water projected in powerful jets
from any direction.
IPX4 – The electrical system is protected
against water splashed against the
components from any direction.
Insulation
Class II equipment
class
Type B
Applied Part complying with the specified
Applied Part
requirements for protection against electrical
shock according to IEC60601-1.
Sound level
45 to 50 dB (A)
The bed has no isolator (main switch). If the bed
needs to be electrically disconnected, it has to be
unplugged from the mains socket.