Electrical Data; Environmental Conditions - Invacare Medley Ergo User Manual

Bed
Hide thumbs Also See for Medley Ergo:
Table of Contents

Advertisement

Invacare® Medley® Ergo
1)
Mattress support mounted in high position.
2)
Mattress support mounted in low position.

9.5 Electrical data

Voltage
Uin = 230 Volt, AC, 50/60 Hz (AC = Alternating
supply
current)
Maximum
Iin max. = 1.5 A
current
input
Intermit-
Int = 10%, max. 2 min / 18 min
tent (peri-
odic motor
operation)
CLASS II equipment
Insulation
class
Type B
Applied Part complying with the specified
Applied
requirements for protection against electrical
Part
shock according to IEC60601-1.
Sound
45 to 50 dB (A)
level
Degree of
The electrical system is IPX4 – protected
protection
against water splashed against the
components from any direction.
Height adjustment motors are IPX4.
The mattress platform motors, control
unit and hand control are IPX6.
IPX6 – These electrical components
are protected against water projected
in powerful jets from any direction.
26
The bed has no isolator (main switch). If the bed
needs to be electrically disconnected, it has to be
unplugged from the mains socket.

9.6 Environmental conditions

Storage and
Transportation
Temperature
-10°C to +50°C
Relative humidity
20% to 80% - not condensing
Atmospheric
pressure
Allow the product to reach operation temperature
before usage:
Warming up from minimum storage temperature
might take up to 24 hours.
Cooling down from maximum storage
temperature might take up to 24 hours.
Operation
+5°C to +40°C
700 hPa to 1060 hPa
60128481-C