Products Solutions Services BA01063D/06/EN/05.16 71340677 Valid as of version 02.00.zz (Device firmware) Operating Instructions Proline Promag W 400 HART Electromagnetic flowmeter...
Page 2
• The manufacturer reserves the right to modify technical data without prior notice. Your Endress+Hauser Sales Center will supply you with current information and updates to these instructions. Endress+Hauser...
Proline Promag W 400 HART Table of contents Table of contents Document information ....6 Installation ..... . . 20 Document function .
Page 4
Table of contents Proline Promag W 400 HART 8.3.8 Calling up help text ....65 10.5.6 Using parameters for device 8.3.9 Changing the parameters ..
Page 5
13.2 Measuring and test equipment ... 141 13.3 Endress+Hauser services ....141 Repairs ......142 14.1 General notes .
Document information Proline Promag W 400 HART Document information Document function These Operating Instructions contain all the information that is required in various phases of the life cycle of the device: from product identification, incoming acceptance and storage, to mounting, connection, operation and commissioning through to troubleshooting, maintenance and disposal.
Proline Promag W 400 HART Document information 1.2.3 Tool symbols Symbol Meaning Torx screwdriver Phillips head screwdriver Open-ended wrench 1.2.4 Symbols for certain types of information Symbol Meaning Permitted Procedures, processes or actions that are permitted. Preferred Procedures, processes or actions that are preferred.
• The W@M Device Viewer : Enter the serial number from the nameplate (www.endress.com/deviceviewer) • The Endress+Hauser Operations App: Enter the serial number from the nameplate or scan the 2-D matrix code (QR code) on the nameplate. For a detailed list of the individual documents along with the documentation code →...
Proline Promag W 400 HART Basic safety instructions Basic safety instructions Requirements for the personnel The personnel for installation, commissioning, diagnostics and maintenance must fulfill the following requirements: ‣ Trained, qualified specialists must have a relevant qualification for this specific function and task.
Verification for borderline cases: ‣ For special fluids and fluids for cleaning, Endress+Hauser is glad to provide assistance in verifying the corrosion resistance of fluid-wetted materials, but does not accept any warranty or liability as minute changes in the temperature, concentration or level of contamination in the process can alter the corrosion resistance properties.
Proline Promag W 400 HART Basic safety instructions It meets general safety standards and legal requirements. It also complies with the EU directives listed in the device-specific EU Declaration of Conformity. Endress+Hauser confirms this by affixing the CE mark to the device. IT security We only provide a warranty if the device is installed and used as described in the Operating Instructions.
Basic safety instructions Proline Promag W 400 HART General notes on the use of passwords • The access code and network key supplied with the device should be changed during commissioning. • Follow the general rules for generating a secure password when defining and managing the access code or network key.
Proline Promag W 400 HART Product description Product description The device consists of a transmitter and a sensor. Two device versions are available: • Compact version – transmitter and sensor form a mechanical unit. • Remote version - transmitter and sensor are mounted in separate locations.
A0028673 and documents present? • If one of the conditions is not satisfied, contact your Endress+Hauser Sales Center. • Depending on the device version, the CD-ROM might not be part of the delivery! The Technical Documentation is available via the Internet or via the Endress+Hauser Operations App, see the "Product identification"...
"Supplementary device-dependent documentation" → 8 • The W@M Device Viewer: Enter the serial number from the nameplate (www.endress.com/deviceviewer) • The Endress+Hauser Operations App: Enter the serial number from the nameplate or scan the 2-D matrix code (QR code) on the nameplate. 4.2.1...
Incoming acceptance and product identification Proline Promag W 400 HART 4.2.2 Sensor nameplate Order Code: Ser.No.: Ext. ord. cd.: Material: Patents Date: A0032085 3 Example of sensor nameplate Name of the sensor Manufacturing location Order code Serial number (ser. no.) Extended order code (Ext.
Page 17
Proline Promag W 400 HART Incoming acceptance and product identification 4.2.3 Symbols on measuring device Symbol Meaning WARNING! This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in serious or fatal injury. Reference to documentation Refers to the corresponding device documentation.
Storage and transport Proline Promag W 400 HART Storage and transport Storage conditions Observe the following notes for storage: • Store in the original packaging to ensure protection from shock. • Do not remove protective covers or protective caps installed on process connections.
Proline Promag W 400 HART Storage and transport 5.2.2 Measuring devices with lifting lugs CAUTION Special transportation instructions for devices with lifting lugs ‣ Only use the lifting lugs fitted on the device or flanges to transport the device. ‣...
Installation Proline Promag W 400 HART Installation Installation conditions 6.1.1 Mounting position Mounting location A0029343 Preferably install the sensor in an ascending pipe, and ensure a sufficient distance to the next pipe elbow: h ≥ 2 × DN Not necessary in the case of order code for "Design", option C "Insertion length short ISO/DVGW to DN300, without inlet/outlet runs, constricted meas.tube"...
Page 21
Proline Promag W 400 HART Installation Installation in partially filled pipes A partially filled pipe with a gradient necessitates a drain-type configuration. A0029257 For heavy sensors DN ≥ 350 (14") A0016276 Orientation The direction of the arrow on the sensor nameplate helps you to install the sensor according to the flow direction (direction of medium flow through the piping).
Page 22
Installation Proline Promag W 400 HART Horizontal A0029344 EPD electrode for empty pipe detection Measuring electrodes for signal detection Reference electrode for potential equalization • Ideally, the measuring electrode plane should be horizontal. This prevents brief insulation of the two measuring electrodes by entrained air bubbles.
• Protect the display against impact. • Protect the display from abrasion by sand in desert areas. You can order a display guard from Endress+Hauser : → 144 Temperature tables Observe the interdependencies between the permitted ambient and fluid temperatures when operating the device in hazardous areas.
Page 24
Installation Proline Promag W 400 HART A0029004 8 Measures to avoid device vibrations (L > 10 m (33 ft)) Adapters Suitable adapters to DIN EN 545 (double-flange reducers) can be used to install the sensor in larger-diameter pipes. The resultant increase in the rate of flow improves measuring accuracy with very slow-moving fluids.
Proline Promag W 400 HART Installation [ S/cm] µ L max L max A0016539 9 Permitted length of connecting cable for remote version = permitted range Colored area = length of connecting cable in [m] ([ft]) = fluid conductivity [µS/cm]...
Installation Proline Promag W 400 HART protective measures. The device is mounted in accordance with the usual regional installation regulations (e.g. EN DIN 1610). A0029321 Mounting the measuring device 6.2.1 Required tools For transmitter • Torque wrench • For wall mounting: Open-ended wrench for hexagonal screw max.
Page 27
Proline Promag W 400 HART Installation 3. If using ground disks, comply with the Installation Instructions provided. 4. Observe required screw tightening torques → 27. 5. Install the measuring device or turn the transmitter housing so that the cable entries do not point upwards.
Installation Proline Promag W 400 HART 6.2.5 Turning the transmitter housing To provide easier access to the connection compartment or display module, the transmitter housing can be turned. TX 20 A0032086 1. Loosen the fixing screws of the housing cover (when reassembling, pay attention to the tightening torque →...
Page 35
Proline Promag W 400 HART Installation TX 20 A0032088 7. Loosen the fixing screws of the main electronics module (when reassembling, pay attention to the tightening torque → 35). 8. Remove the main electronics module. 4 mm A0032089 9. Loosen the fixing screws of the transmitter housing (when reassembling, pay attention to the tightening torque →...
Installation Proline Promag W 400 HART NOTICE Plug of the smart sensor electronics module connected incorrectly! No measuring signal is output. ‣ Plug in the plug of the smart sensor electronics module as per the coding. A0021585 ‣ Reverse the procedure to reassemble the measuring device.
Proline Promag W 400 HART Installation Post-installation check Is the device undamaged (visual inspection)? Does the measuring device conform to the measuring point specifications? For example: • Process temperature • Process pressure (refer to the section on "Pressure-temperature ratings" in the "Technical Information"...
Electrical connection Proline Promag W 400 HART Electrical connection NOTICE The measuring device does not have an internal circuit breaker. ‣ For this reason, assign the measuring device a switch or power-circuit breaker so that the power supply line can be easily disconnected from the mains.
Page 39
Proline Promag W 400 HART Electrical connection Coil current cable Standard cable (18 AWG) with common, braided copper shield ( ~9 mm 3 ×0.75 mm (0.35 in)) Conductor resistance ≤37 Ω/km (0.011 Ω/ft) Capacitance: core/core, ≤120 pF/m (37 pF/ft) shield grounded Operating temperature –20 to +80 °C (–68 to +176 °F)
Electrical connection Proline Promag W 400 HART 7.1.2 Required tools • Torque wrench • For cable entries: Use corresponding tools • Wire stripper • When using stranded cables: Crimper for wire end ferrule 7.1.3 Terminal assignment Transmitter The sensor can be ordered with terminals.
Electrical connection Proline Promag W 400 HART 2. If the measuring device is supplied without cable glands: Provide suitable cable gland for corresponding connecting cable. 3. If the measuring device is supplied with cable glands: Observe requirements for connecting cables → 38.
Page 44
Electrical connection Proline Promag W 400 HART 3. Connect the transmitter. Connecting the connecting cable to the sensor connection housing 42 41 A0032103 17 Sensor: connection module 1. Loosen the securing clamp of the housing cover. 2. Unscrew and lift off the housing cover.
Proline Promag W 400 HART Electrical connection Connecting the connecting cable to the transmitter TX 20 6 5 7 8 4 37 36 42 41 A0032102 18 Transmitter: main electronics module with terminals 1. Loosen the 4 fixing screws on the housing cover.
Electrical connection Proline Promag W 400 HART TX 20 + - + - + - + - 23 20 21 26 27 24 25 10 (0.4) A0032104 19 Connecting the supply voltage and 0-20 mA/4-20 mA HART with additional outputs and inputs 1.
Page 47
Proline Promag W 400 HART Electrical connection Connection example, standard scenario Metal, grounded pipe A0016315 20 Potential equalization via measuring tube Connection example in special situations Unlined and ungrounded metal pipe This connection method also applies in situations where: •...
Page 48
Electrical connection Proline Promag W 400 HART A0029339 22 Potential equalization via ground terminal and ground disks 1. Connect the ground disks to the ground terminal via the ground cable. 2. Connect the ground disks to ground potential. For remote device versions, the ground terminal in the example always refers to the sensor and not to the transmitter.
Proline Promag W 400 HART Electrical connection Special connection instructions 7.3.1 Connection examples Current output 4 to 20 mA HART 4...20 mA A0029055 23 Connection example for 4 to 20 mA HART current output (active) Automation system with current input (e.g. PLC) Cable shield: the cable shield must be grounded at both ends to comply with EMC requirements;...
Page 50
Electrical connection Proline Promag W 400 HART Pulse/frequency output 12345 A0028761 25 Connection example for pulse/frequency output (passive) Automation system with pulse/frequency input (e.g. PLC) Power supply Transmitter: Observe input values → 152 Switch output A0028760 26 Connection example for switch output (passive) Automation system with switch input (e.g.
Proline Promag W 400 HART Electrical connection Ensuring the degree of protection 7.4.1 Degree of protection IP66/67, Type 4X enclosure The measuring device fulfills all the requirements for the IP66/67 degree of protection, Type 4X enclosure. To guarantee IP66/67 degree of protection, Type 4X enclosure, carry out the following steps after the electrical connection: 1.
Page 52
Electrical connection Proline Promag W 400 HART Only for remote version: is the sensor connected to the right transmitter? Check the serial number on the nameplate of the sensor and transmitter. Does the supply voltage match the specifications on the transmitter nameplate → 41? ...
Proline Promag W 400 HART Operation options Operation options Overview of operation options A0029295 Local operation via display module Computer with Web browser (e.g. Internet Explorer) or with operating tool (e.g. FieldCare, DeviceCare, AMS Device Manager, SIMATIC PDM) Field Xpert SFX350 or SFX370...
Operation options Proline Promag W 400 HART Structure and function of the operating menu 8.2.1 Structure of the operating menu For an overview of the operating menu for experts: "Description of Device Parameters" document supplied with the device→ 191...
Proline Promag W 400 HART Operation options 8.2.2 Operating philosophy The individual parts of the operating menu are assigned to certain user roles (operator, maintenance etc.). Each user role contains typical tasks within the device lifecycle. For custody transfer, once the device has been put into circulation or sealed, its operation is restricted.
Operation options Proline Promag W 400 HART Menu/parameter User role and tasks Content/meaning Expert function-oriented Tasks that require detailed Contains all the parameters of the device and makes it possible to access knowledge of the function of the these parameters directly using an access code. The structure of this menu is...
Page 57
Proline Promag W 400 HART Operation options Display area In the display area, each measured value is prefaced by certain symbol types for further description: Measured variable Measurement channel Diagnostic behavior number ↓ ↓ ↓ Example Appears only if a diagnostics event is present for this measured variable.
Operation options Proline Promag W 400 HART 8.3.2 Navigation view In the submenu In the wizard /../Operation 0091-1 /../Curr. output 1 Access stat.disp Assign curr. Operator Volume flow Locking status Display A0013993-EN A0016327-EN Navigation view Navigation path to current position...
Page 59
Proline Promag W 400 HART Operation options Display area Menus Symbol Meaning Operation Appears: • In the menu next to the "Operation" selection • At the left in the navigation path in the Operation menu Setup Appears: • In the menu next to the "Setup" selection •...
Proline Promag W 400 HART Operation options Selection of letters from a to z. abc _ … Selection of special characters. "'^ _ … ~& Confirms selection. Switches to the selection of the correction tools. Exits the input without applying the changes.
Operation options Proline Promag W 400 HART Meaning Enter key For operational display • Pressing the key briefly opens the operating menu. • Pressing the key for 2 s opens the context menu. In a menu, submenu • Pressing the key briefly: –...
Page 63
Proline Promag W 400 HART Operation options 2. Press + simultaneously. The context menu is closed and the operational display appears. Calling up the menu via the context menu 1. Open the context menu. 2. Press to navigate to the desired menu.
Operation options Proline Promag W 400 HART 8.3.6 Navigating and selecting from list Different operating elements are used to navigate through the operating menu. The navigation path is displayed on the left in the header. Icons are displayed in front of the individual menus.
Proline Promag W 400 HART Operation options The direct access code consists of a 4-digit number and the channel number, which identifies the channel of a process variable: e.g. 0914-1. In the navigation view, this appears on the right-hand side in the header of the selected parameter.
Operation options Proline Promag W 400 HART 8.3.9 Changing the parameters For a description of the editing display - consisting of text editor and numeric editor - with symbols → 60, for a description of the operating elements → 61 Example: Changing the tag name in the "Tag description"...
Proline Promag W 400 HART Operation options Ent. access code Invalid or out of range input value Min:0 Max:9999 A0014049-EN 8.3.10 User roles and related access authorization The two user roles "Operator" and "Maintenance" have different write access to the parameters if the customer defines a user-specific access code.
Operation options Proline Promag W 400 HART Local operation with touch control The keypad lock is switched on and off via the context menu. Switching on the keypad lock The keypad lock is switched on automatically: • Each time the device is restarted.
Page 69
Proline Promag W 400 HART Operation options Computer software Software Interface CDI-RJ45 WLAN Recommended operating • Microsoft Windows 7 or higher. systems • Mobile operating systems: – iOS – Android Microsoft Windows XP is supported. Web browsers supported • Microsoft Internet Explorer 8 or higher •...
Operation options Proline Promag W 400 HART 8.4.3 Establishing a connection Via service interface (CDI-RJ45) Preparing the measuring device Configuring the Internet protocol of the computer The following information refers to the default Ethernet settings of the device. IP address of the device: 192.168.1.212 (factory setting) 1.
Proline Promag W 400 HART Operation options Disconnecting ‣ After configuring the device: Terminate the WLAN connection between the operating unit and measuring device. Starting the Web browser 1. Start the Web browser on the computer. 2. Enter the IP address of the Web server in the address line of the Web browser: 192.168.1.212...
Operation options Proline Promag W 400 HART 8.4.5 User interface A0029418 Function row Local display language Navigation area Header The following information appears in the header: • Device tag • Device status with status signal → 128 • Current measured values...
Proline Promag W 400 HART Operation options Working area Depending on the selected function and the related submenus, various actions can be performed in this area: • Configuring parameters • Reading measured values • Calling up help text • Starting an upload/download 8.4.6...
Operation options Proline Promag W 400 HART 8.5.1 Connecting the operating tool Via service interface (CDI-RJ45) open press A0029163 30 Connection via service interface (CDI-RJ45) Computer with Web browser (e.g. Microsoft Internet Explorer, Microsoft Edge) for accessing the integrated device Web server or with "FieldCare", "DeviceCare"...
Proline Promag W 400 HART Operation options Antenna type • Internal • External (optional) Only one channel active in each case! Materials • External antenna: ASA (acrylic ester-styrene-acrylonitrile) plastic and nickel-plated brass • Adapter: stainless steel and copper Configuring the Internet protocol of the mobile terminal NOTICE If the WLAN connection is lost during the configuration, settings made may be lost.
FieldCare Function scope FDT-based plant asset management tool from Endress+Hauser. It can configure all smart field devices in a system and helps you manage them. By using the status information, it is also a simple but effective way of checking their status and condition.
DeviceCare Function scope Tool to connect and configure Endress+Hauser field devices. The fastest way to configure Endress+Hauser field devices is with the dedicated "DeviceCare" tool. Together with the device type managers (DTMs) it presents a convenient, comprehensive solution. For details, see Innovation Brochure IN01047S Source for device description files See information →...
Operation options Proline Promag W 400 HART 8.5.6 SIMATIC PDM Function scope SIMATIC PDM is a standardized, manufacturer-independent program from Siemens for the operation, configuration, maintenance and diagnosis of intelligent field devices via HART protocol. Source for device description files See data →...
Proline Promag W 400 HART System integration System integration Overview of device description files 9.1.1 Current version data for the device Firmware version 02.00.zz • On the title page of the Operating instructions • On the transmitter nameplate • Firmware version Diagnostics →...
Page 80
System integration Proline Promag W 400 HART Dynamic variables Measured values (HART device variables) Primary dynamic variable (PV) Volume flow Secondary dynamic variable (SV) Totalizer 1 Tertiary dynamic variable (TV) Totalizer 2 Quaternary dynamic variable (QV) Totalizer 3 The assignment of the measured variables to the dynamic variables can be modified and...
Proline Promag W 400 HART System integration Other settings Burst mode functionality in accordance with HART 7 Specification: Navigation "Expert" menu → Communication → HART output → Burst configuration → Burst configuration 1 to n ‣ Burst configuration ‣ Burst configuration 1 to n →...
Page 82
System integration Proline Promag W 400 HART Parameter Description Selection / User entry Factory setting Burst variable 0 For HART command 9 and 33: select the • Volume flow Volume flow HART device variable or the process variable. • Mass flow • Corrected volume flow •...
Proline Promag W 400 HART Commissioning Commissioning 10.1 Function check Before commissioning the measuring device: ‣ Make sure that the post-installation and post-connection checks have been performed. • "Post-installation check" checklist → 37 • "Post-connection check" checklist → 51 10.2...
Commissioning Proline Promag W 400 HART X X X X X X X 20.50 0104-1 Main menu Display language English Display/operat. Setup Main menu Display/operat. Setup Diagnostic / ../Setup Medium selection XXXXXXXXX XXXXXXXXX A0032222-EN 32 Taking the example of the local display Navigation "Setup"...
Proline Promag W 400 HART Commissioning XXXXXXXXX A0029422 33 Header of the operational display with tag name Tag name Enter the tag name in the "FieldCare" operating tool → 77 Navigation "Setup" menu → Device tag Parameter overview with brief description...
Commissioning Proline Promag W 400 HART Parameter overview with brief description Parameter Prerequisite Description Selection Factory setting Volume flow unit – Select volume flow unit. Unit choose list Country-specific: • l/h Result • gal/min (us) The selected unit applies for: •...
Page 87
Proline Promag W 400 HART Commissioning Active level → 87 Response time status input → 87 Parameter overview with brief description Parameter Description Selection / User entry Factory setting Assign status input Select function for the status input. • Off •...
Commissioning Proline Promag W 400 HART 10.4.4 Configuring the current output The Current output wizard guides you systematically through all the parameters that have to be set for configuring the current output. Navigation "Setup" menu → Current output 1 ‣...
Proline Promag W 400 HART Commissioning Parameter Prerequisite Description Selection / User Factory setting entry 20 mA value One of the following options is Enter 20 mA value. Signed floating-point Depends on country selected in the Current span number and nominal parameter (→ 88): diameter •...
Page 90
Commissioning Proline Promag W 400 HART Failure mode → 90 Invert output signal → 90 Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry Operating mode – Define the output as a pulse, • Pulse Pulse frequency or switch output.
Page 91
Proline Promag W 400 HART Commissioning Maximum frequency value → 91 Measuring value at minimum → 92 frequency Measuring value at maximum → 92 frequency → 92 Failure mode → 92 Failure frequency Invert output signal →...
Page 92
Commissioning Proline Promag W 400 HART Parameter Prerequisite Description Selection / User Factory setting entry Measuring value at minimum In the Operating mode Enter measured value for Signed floating-point Depends on country frequency parameter the Frequency minmum frequency. number and nominal option is selected and in the...
Page 93
Proline Promag W 400 HART Commissioning Configuring the switch output Navigation "Setup" menu → Pulse/frequency/switch output 1 to n ‣ Pulse/frequency/switch output 1 to n Operating mode → 93 Switch output function → 93 Assign diagnostic behavior → 93 →...
Page 94
Commissioning Proline Promag W 400 HART Parameter Prerequisite Description Selection / User Factory setting entry Assign limit • In the Operating mode Select process variable for limit • Off Volume flow parameter, the Switch function. • Volume flow option is selected. • Mass flow •...
Proline Promag W 400 HART Commissioning 10.4.6 Configuring the local display The Display wizard guides you systematically through all the parameters that can configured for configuring the local display. Navigation "Setup" menu → Display ‣ Display → 95 Format display →...
Commissioning Proline Promag W 400 HART Parameter Prerequisite Description Selection / User Factory setting entry 100% bargraph value 1 A local display is provided. Enter 100% value for bar Signed floating-point Depends on country graph display. number and nominal diameter Value 2 display A local display is provided.
Page 97
Proline Promag W 400 HART Commissioning Measuring mode output 1 to n → 97 Assign pulse output 1 to n → 98 Measuring mode output 1 to n → 98 Parameter overview with brief description Parameter Prerequisite Description User entry / ...
Commissioning Proline Promag W 400 HART Parameter Prerequisite Description User entry / Factory setting Selection Assign pulse output In the Operating mode Select process variable for • Off parameter, the Pulse option is pulse output. • Volume flow selected. • Mass flow •...
Page 99
Proline Promag W 400 HART Commissioning Parameter Prerequisite Description Selection / User Factory setting entry Off value low flow cutoff One of the following options is Enter off value for low flow cut 0 to 100.0 % 50 % selected in the Assign process off.
Commissioning Proline Promag W 400 HART 10.4.9 Configuring empty pipe detection The Empty pipe detection wizard guides you systematically through all the parameters that have to be set for configuring empty pipe detection. Navigation "Setup" menu → Empty pipe detection ‣...
Proline Promag W 400 HART Commissioning 10.5 Advanced settings The Advanced setup submenu together with its submenus contains parameters for specific settings. Navigation to the "Advanced setup" submenu X X X X X X X 20.50 0104-1 Main menu Display language English Display/operat.
Commissioning Proline Promag W 400 HART 10.5.3 Carrying out additional display configurations In the Display submenu you can set all the parameters associated with the configuration of the local display. Navigation "Setup" menu → Advanced setup → Display ‣ Display →...
Page 105
Proline Promag W 400 HART Commissioning Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry Format display A local display is provided. Select how measured values • 1 value, max. size 1 value, max. size are shown on the display.
Page 106
Commissioning Proline Promag W 400 HART Parameter Prerequisite Description Selection / User Factory setting entry 100% bargraph value 3 A selection was made in the Enter 100% value for bar Signed floating-point Value 3 display parameter. graph display. number Decimal places 3...
Proline Promag W 400 HART Commissioning 10.5.4 Performing electrode cleaning The Electrode cleaning circuit wizard guides the user systematically through all the parameters that have to be set for configuring electrode cleaning. The wizard only appears if the device was ordered with an electrode cleaning circuit.
Commissioning Proline Promag W 400 HART Parameter overview with brief description Parameter Description User interface / User entry Factory setting Operating time Indicates how long the device has been in Days (d), hours (h), minutes – operation. (m) and seconds (s) Reset access code Reset access code to factory settings.
Page 111
Proline Promag W 400 HART Commissioning Frequency output simulation 1 to n → 112 Frequency value 1 to n → 112 Pulse output simulation 1 to n → 112 Pulse value 1 to n → 112 Switch output simulation 1 to n →...
Commissioning Proline Promag W 400 HART Parameter Prerequisite Description Selection / User Factory setting entry Value current output 1 In the Current output Enter the current value for 3.59 to 22.5 mA 3.59 mA simulation parameter, the On simulation. option is selected.
Proline Promag W 400 HART Commissioning Defining the access code via local display 1. Navigate to the Define access code parameter (→ 109). 2. Define a max. 16-digit character string comprising numbers, letters and special characters as the access code.
Page 114
Commissioning Proline Promag W 400 HART The parameter values are now read only and cannot be edited any more (exception "Contrast display" parameter): • Via local display • Via service interface (CDI-RJ45) • Via HART protocol + - + -...
Proline Promag W 400 HART Operation Operation 11.1 Reading the device locking status Device active write protection: Locking status parameter Operation → Locking status Function scope of the "Locking status" parameter Options Description None The access status displayed in the Access status display parameter applies →...
Proline Promag W 400 HART Operation 11.4.2 "Totalizer" submenu The Totalizer submenu contains all the parameters needed to display the current measured values for every totalizer. Navigation "Diagnostics" menu → Measured values → Totalizer ‣ Totalizer → 117 Totalizer value 1 to n →...
Operation Proline Promag W 400 HART 11.4.4 Output values The Output values submenu contains all the parameters needed to display the current measured values for every output. The parameters displayed depend on: • The selected device order • The set operating mode of the pulse/frequency/switch outputs Navigation "Diagnostics"...
Proline Promag W 400 HART Operation 11.6 Performing a totalizer reset The totalizers are reset in the Operation submenu: • Control Totalizer • Reset all totalizers Navigation "Operation" menu → Totalizer handling ‣ Totalizer handling Control Totalizer 1 to n →...
Operation Proline Promag W 400 HART 11.6.2 Function scope of the "Reset all totalizers" parameter Options Description Cancel No action is executed and the user exits the parameter. Reset + totalize Resets all totalizers to 0 and restarts the totaling process. This deletes all the flow values previously totalized.
Page 121
Proline Promag W 400 HART Operation Data logging control → 121 Data logging status → 121 Entire logging duration → 121 Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry / User interface Assign channel 1 to n...
Diagnostics and troubleshooting Proline Promag W 400 HART Diagnostics and troubleshooting 12.1 General troubleshooting For local display Error Possible causes Solution Local display dark and no output Supply voltage does not match the Apply the correct supply voltage signals value indicated on the nameplate.
Page 123
Proline Promag W 400 HART Diagnostics and troubleshooting For access Error Possible causes Solution No write access to parameters Hardware write protection enabled Set the write protection switch on main electronics module to the OFF position → 113. No write access to parameters Current user role has limited access 1.
Diagnostics and troubleshooting Proline Promag W 400 HART Error Possible causes Solution No or incomplete display of • JavaScript not enabled 1. Enable JavaScript. contents in the Web browser • JavaScript cannot be enabled 2. Enter http://192.168.1.212/ basic.html as the IP address.
Proline Promag W 400 HART Diagnostics and troubleshooting 12.3 Diagnostic information on local display 12.3.1 Diagnostic message Faults detected by the self-monitoring system of the measuring device are displayed as a diagnostic message in alternation with the operational display. Operational display in alarm condition...
Page 126
Diagnostics and troubleshooting Proline Promag W 400 HART Diagnostic behavior Symbol Meaning Alarm • Measurement is interrupted. • Signal outputs and totalizers assume the defined alarm condition. • A diagnostic message is generated. • The background lighting changes to red.
Proline Promag W 400 HART Diagnostics and troubleshooting 12.3.2 Calling up remedial measures X X X X X X X X X X X X X X 20.50 S801 Supply voltage Menu Diagnostic list Diagnostics 1 S801 Supply voltage Diagnostics 2...
Diagnostics and troubleshooting Proline Promag W 400 HART A0031056 Status area with status signal Diagnostic information → 126 Remedy information with Service ID In addition, diagnostic events which have occurred can be shown in the Diagnostics menu: • Via parameter •...
Proline Promag W 400 HART Diagnostics and troubleshooting 12.5 Diagnostic information in DeviceCare or FieldCare 12.5.1 Diagnostic options Any faults detected by the measuring device are displayed on the home page of the operating tool once the connection has been established.
Diagnostics and troubleshooting Proline Promag W 400 HART 12.5.2 Calling up remedy information Remedy information is provided for every diagnostic event to ensure that problems can be rectified quickly: • On the home page Remedy information is displayed in a separate field below the diagnostics information.
Proline Promag W 400 HART Diagnostics and troubleshooting Available status signals Configuration as per HART 7 Specification (Condensed Status), in accordance with NAMUR NE107. Symbol Meaning Failure A device error is present. The measured value is no longer valid. A0013956 Function check The device is in service mode (e.g.
Page 132
Diagnostics and troubleshooting Proline Promag W 400 HART Diagnostic Short text Remedy instructions Status Diagnostic number signal behavior [from the [from the factory] factory] Software incompatible 1. Check software Alarm 2. Flash or change main electronics module Modules incompatible 1. Check electronic modules Alarm 2.
Page 133
Proline Promag W 400 HART Diagnostics and troubleshooting Diagnostic Short text Remedy instructions Status Diagnostic number signal behavior [from the [from the factory] factory] Diagnostic of configuration Data transfer 1. Check connection Alarm 2. Retry data transfer Processing download Download active, please wait...
Diagnostics and troubleshooting Proline Promag W 400 HART Diagnostic Short text Remedy instructions Status Diagnostic number signal behavior [from the [from the factory] factory] Diagnostic of process Current loop 1. Check wiring Alarm 2. Change I/O module Electronic temperature Reduce ambient temperature...
Proline Promag W 400 HART Diagnostics and troubleshooting Operating time from restart → 135 Operating time → 135 Parameter overview with brief description Parameter Prerequisite Description User interface Actual diagnostics A diagnostic event has occurred. Shows the current occured diagnostic...
Diagnostics and troubleshooting Proline Promag W 400 HART Navigation path Diagnostics menu → Event logbook submenu → Event list / ../Eventlist I1091 Config. change I1157 Mem.err. ev.list 0d01h19m10s F311 Electr. failure A0014008-EN 38 Taking the example of the local display •...
Page 137
Proline Promag W 400 HART Diagnostics and troubleshooting Info number Info name I1089 Power on I1090 Configuration reset I1091 Configuration changed I1092 Embedded HistoROM deleted I1137 Electronic changed I1151 History reset I1155 Reset electronic temperature I1156 Memory error trend I1157...
Diagnostics and troubleshooting Proline Promag W 400 HART Info number Info name I1643 Custody transfer logbook cleared I1649 Hardware write protection activated I1650 Hardware write protection deactivated I1651 Custody transfer parameter changed I1725 Sensor electronic module (ISEM) changed 12.11 Resetting the measuring device Using theDevice reset parameter (→...
Page 139
Proline Promag W 400 HART Diagnostics and troubleshooting ENP version → 139 Device revision → 139 Device ID → 139 Device type → 140 Manufacturer ID → 140 Parameter overview with brief description Parameter Description...
"Manufacturer' s information" document. The manufacturer' s information is available: • In the Download Area of the Endress+Hauser web site: www.endress.com → Downloads • Specify the following details: –...
Replacement seals (accessory) → 190 13.2 Measuring and test equipment Endress+Hauser offers a wide variety of measuring and test equipment, such as W@M or device tests. Your Endress+Hauser Sales Center can provide detailed information on the services. List of some of the measuring and testing equipment: → 144 13.3...
• The measuring devices have a modular design. • Spare parts are grouped into logical kits with the associated Installation Instructions. • Repairs are carried out by Endress+Hauser Service or by appropriately trained customers. • Certified devices can only be converted to other certified devices by Endress+Hauser Service or at the factory.
Proline Promag W 400 HART Repairs 14.5 Disposal 14.5.1 Removing the measuring device 1. Switch off the device. WARNING Danger to persons from process conditions. ‣ Beware of hazardous process conditions such as pressure in the measuring device, high temperatures or aggressive fluids.
Various accessories, which can be ordered with the device or subsequently from Endress +Hauser, are available for the device. Detailed information on the order code in question is available from your local Endress+Hauser sales center or on the product page of the Endress+Hauser website: www.endress.com.
Accessories Description Applicator Software for selecting and sizing Endress+Hauser measuring devices: • Choice of measuring devices for industrial requirements • Calculation of all the necessary data for identifying the optimum flowmeter: e.g. nominal diameter, pressure loss, flow velocity and accuracy.
Accessories Proline Promag W 400 HART 15.4 System components Accessories Description Memograph M graphic The Memograph M graphic display recorder provides information on all relevant display recorder measured variables. Measured values are recorded correctly, limit values are monitored and measuring points analyzed. The data are stored in the 256 MB internal memory and also on a SD card or USB stick.
Proline Promag W 400 HART Technical data Technical data 16.1 Application The measuring device described in this manual is intended only for flow measurement of liquids with a minimum conductivity of 5 μS/cm. Depending on the version ordered, the measuring device can also measure potentially explosive, flammable, poisonous and oxidizing media.
Page 148
Technical data Proline Promag W 400 HART Nominal Recommended Factory settings diameter flow Full scale value current min./max. full scale value Pulse value Low flow cut off output (v ~ 0.3/10 m/s) (~ 2 pulse/s) (v ~ 0.04 m/s) (v ~ 2.5 m/s)
Page 149
Proline Promag W 400 HART Technical data Flow characteristic values in SI units Nominal Recommended Factory settings diameter flow Full scale value current min./max. full scale value Pulse value Low flow cut off output (v ~ 0.12/5 m/s) (~ 4 pulse/s) (v ~ 0.01 m/s)
Page 150
Technical data Proline Promag W 400 HART Nominal Recommended Factory settings diameter flow Full scale value current min./max. full scale value Pulse value Low flow cut off output (v ~ 0.3/10 m/s) (~ 2 pulse/s) (v ~ 0.04 m/s) (v ~ 2.5 m/s)
Input signal External measured values Various pressure transmitters and temperature measuring devices can be ordered from Endress+Hauser: see "Accessories" section → 146 It is recommended to read in external measured values to calculate the following measured variables: Corrected volume flow...
Page 152
Technical data Proline Promag W 400 HART Pulse/frequency/switch output Function • With the order code for "Output; Input", option H: output 2 can be set as a pulse or frequency output • With the order code for "Output; Input", option I: output 2 and 3 can be set as a pulse, frequency or switch output •...
Page 153
Proline Promag W 400 HART Technical data Current output 4 to 20 mA 4 to 20 mA Failure mode Choose from: • 4 to 20 mA in accordance with NAMUR recommendation NE 43 • 4 to 20 mA in accordance with US •...
Technical data Proline Promag W 400 HART Web server Plain text display With information on cause and remedial measures Light emitting diodes (LED) Status information Status indicated by various light emitting diodes The following information is displayed depending on the device version: •...
Proline Promag W 400 HART Technical data Current consumption Transmitter Order code for "Power supply" Maximum Maximum Current consumption switch-on current Option L: AC 100 to 240 V 145 mA 25 A (< 5 ms) Option L: AC/DC 24 V 350 mA 27 A (<...
Page 156
Technical data Proline Promag W 400 HART Volume flow • ±0.5 % o.r. ± 1 mm/s (0.04 in/s) • Optional: ±0.2 % o.r. ± 2 mm/s (0.08 in/s) Fluctuations in the supply voltage do not have any effect within the specified range.
Proline Promag W 400 HART Technical data Flat Spec flow values0.2 % Nominal diameter [mm] [in] [m/s] [ft/s] [m/s] [ft/s] 25 to 600 1 to 24 4.92 50 to 300 2 to 12 1.97 Order code for "Design", option A "Insertion length short, ISO/DVGW to DN400, DN450-2000 1:1" and order code for "Design", option B "Insertion length long, ISO/DVGW to DN400, DN450-2000 1:1.3"...
Atmosphere If a plastic transmitter housing is permanently exposed to certain steam and air mixtures, this can damage the housing. If you are unsure, please contact your Endress+Hauser Sales Center for clarification. Degree of protection Transmitter • As standard: IP66/67, type 4X enclosure •...
Proline Promag W 400 HART Technical data Impact resistance Rough handling shocks according to IEC 60068-2-31 Mechanical load • Protect the transmitter housing against mechanical effects, such as shock or impact; the use of the remote version is sometimes preferable.
Page 160
Technical data Proline Promag W 400 HART Flow limit The diameter of the pipe and the flow rate determine the nominal diameter of the sensor. The optimum flow velocity is between 2 to 3 m/s (6.56 to 9.84 ft/s). Also match the velocity of flow (v) to the physical properties of the fluid: •...
Page 162
Technical data Proline Promag W 400 HART Weight in SI units Standard version Order code for "Design", option A "Insertion length short, ISO/DVGW to DN400, DN450-2000 1:1" and order code for "Design", option B "Insertion length long, ISO/DVGW to DN400, DN450-2000 1:1.3"...
Page 163
Proline Promag W 400 HART Technical data Order code for "Design", option C "Insertion length short ISO/DVGW to DN300, without inlet/ outlet runs, constricted meas.tube" EN 1092-1 (DIN 2501) Pressure rating Weight [kg] [mm] Order code for "Housing", option M, Q Order code for "Housing", option A, R...
Page 164
Technical data Proline Promag W 400 HART JIS B2220, 10K Weight [kg] [mm] Order code for "Housing", option M, Q Order code for "Housing", option A, R Polycarbonate plastic Aluminum, AlSi10Mg, coated 11.7 13.7 19.7 23.7 40.7 68.7 70.7 Order code for "Design", option C "Insertion length short ISO/DVGW to DN300, without inlet/ outlet runs, constricted meas.tube"...
Page 165
Proline Promag W 400 HART Technical data EN 1092-1 (DIN 2501) Order code for "Housing", option M, Q [mm] Polycarbonate plastic Weight [kg] PN 6 PN 10 PN 16 1 000 1 200 1 400 1 115 1 201 1 600...
Page 166
Technical data Proline Promag W 400 HART Order code for "Calibration flow", options H and K, or options H/K or order code for "Sensor option", option CA Option Description MID Type Examination Cert MI-001 OIML R49 Class 2 IP66/67, Type 4X, fully welded; corrosion protection EN ISO 12944 C5-M Order code for "Design", option A "Insertion length short, ISO/DVGW to DN400,...
Page 167
Proline Promag W 400 HART Technical data Order code for "Design", option A "Insertion length short, ISO/DVGW to DN400, DN450-2000 1:1" and order code for "Design", option B "Insertion length long, ISO/DVGW to DN400, DN450-2000 1:1.3" AS 2129, PN 16...
Page 168
Technical data Proline Promag W 400 HART Order code for "Design", option C "Insertion length short ISO/DVGW to DN300, without inlet/ outlet runs, constricted meas.tube" JIS B2220, 10K Weight [kg] [mm] Order code for "Housing", option M, Q Order code for "Housing", option A, R...
Page 169
Proline Promag W 400 HART Technical data Order code for "Design", option C "Insertion length short ISO/DVGW to DN300, without inlet/ outlet runs, constricted meas.tube" ASME B16.5, Class 150 Weight [lbs] [in] Order code for "Housing", option M, Q Order code for "Housing", option A, R...
Page 170
Technical data Proline Promag W 400 HART AWWA C207, Class D Weight [lbs] [in] Order code for "Housing", option M, Q Order code for "Housing", option A, R Polycarbonate plastic Aluminum, AlSi10Mg, coated 590.5 704.5 848.5 1 039 1 040.5 1 297 1 298.5...
Page 171
Proline Promag W 400 HART Technical data ASME B16.5, Class 150 Weight [in] [lbs] 118.2 172.4 236.8 Transmitter remote version Wall-mount housing Depends on the material of the wall-mount housing: • Polycarbonate plastic: 1.3 kg (2.9 lb) • Aluminum, AlSi10Mg, coated: 2.0 kg (4.4 lb)
Page 173
Proline Promag W 400 HART Technical data Order code for "Design", option A "Insertion length short, ISO/DVGW to DN400, DN450-2000 1:1" and order code for "Design", option B "Insertion length long, ISO/DVGW to DN400, DN450-2000 1:1.3" JIS B2220, 10K Weight...
Page 175
Proline Promag W 400 HART Technical data Order code for "Design", option A "Insertion length short, ISO/DVGW to DN400, DN450-2000 1:1" and order code for "Design", option B "Insertion length long, ISO/DVGW to DN400, DN450-2000 1:1.3" EN 1092-1 (DIN 2501)
Page 176
Technical data Proline Promag W 400 HART Order code for "Design", option C "Insertion length short ISO/DVGW to DN300, without inlet/ outlet runs, constricted meas.tube" AS 4087, PN 16 [kg] [mm] 14.4 16.0 30.3 Order code for "Design", option A "Insertion length short, ISO/DVGW to DN400, DN450-2000 1:1"...
Page 177
Proline Promag W 400 HART Technical data Weight in US units Standard version Order code for "Design", option A "Insertion length short, ISO/DVGW to DN400, DN450-2000 1:1" and order code for "Design", option B "Insertion length long, ISO/DVGW to DN400, DN450-2000 1:1.3"...
Page 178
Technical data Proline Promag W 400 HART AWWA C207, Class D Weight [in] [lbs] 2 421 3 083 4 847 5 949 8 154 9 036 10 139 Order code for "Design", option A Option A "Insertion length short; ISO/DVGW to DN400, DN450-2000 1:1"...
Page 179
Proline Promag W 400 HART Technical data Order code for "Sensor option", option CA Option CA "IP66/67, Type 4X, fully welded; corrosion protection EN ISO 12944 C5-M" Order code for "Design", option A "Insertion length short, ISO/DVGW to DN400, DN450-2000 1:1" and order code for "Design", option B "Insertion length long, ISO/DVGW to DN400, DN450-2000 1:1.3"...
Page 180
Technical data Proline Promag W 400 HART Nominal diameter Pressure rating Measuring tube internal diameter EN (DIN) ASME AS 2129 Hard rubber Polyurethane AWWA AS 4087 [mm] [in] [mm] [in] [mm] [in] PN 10 Class 150 Table E, PN 16 8.03...
Page 181
Proline Promag W 400 HART Technical data Compact version, inclined • Order code for "Housing", option R "Compact, aluminum coated": Aluminum, AlSi10Mg, coated • Order code for "Housing", option Q: polycarbonate plastic • Window material: – For order code for "Housing", option R: glass –...
Page 182
Technical data Proline Promag W 400 HART Sensor connection housing • Standard: aluminum, AlSi10Mg, coated (IP66/67) • Option: – Polycarbonate for IP68 with DN 50 to 300 (2 to 12") – Polycarbonate for order code "Sensor option", option CA…CE "Corrosion protection" with DN 350 to 2000 (14 to 78")
Page 183
Proline Promag W 400 HART Technical data ASME B16.5 • DN 25 to 1200 (1 to 48"): Stainless steel, F316L similar to 1.4404 • DN 25 to 300 (1 to 12") Carbon steel, A105 similar to 1.0432 • DN 350 to 1200 (14 to 48")
Technical data Proline Promag W 400 HART Process connections • EN 1092-1 (DIN 2501) – DN ≤ 300: fixed flange (PN 10/16/25/40) = form A – DN ≥ 350: fixed flange (PN 6/10/16/25) = flat face – DN 450 to 2000 : fixed flange (PN 6/10/16) = flat face •...
Page 185
Proline Promag W 400 HART Technical data A0032074 44 Operation with touch control Display elements • 4-line, illuminated, graphic display • White background lighting; switches to red in event of device errors • Format for displaying measured variables and status variables can be individually configured •...
Page 186
Technical data Proline Promag W 400 HART Supported operating Operating unit Interface Additional information tools FieldCare SFE500 Notebook, PC or tablet • CDI-RJ45 service → 145 with Microsoft Windows interface system • WLAN interface • Fieldbus protocol Device Xpert...
Page 187
Proline Promag W 400 HART Technical data HistoROM The measuring device features HistoROM data management. HistoROM data management data management comprises both the storage and import/export of key device and process data, making operation and servicing far more reliable, secure and efficient.
EU Directives. These are listed in the corresponding EU Declaration of Conformity along with the standards applied. Endress+Hauser confirms successful testing of the device by affixing to it the CE mark. C-Tick symbol The measuring system meets the EMC requirements of the "Australian Communications and Media Authority (ACMA)".
The application packages can be ordered with the device or subsequently from Endress+Hauser. Detailed information on the order code in question is available from your local Endress+Hauser sales center or on the product page of the Endress+Hauser website: www.endress.com.
• The W@M Device Viewer : Enter the serial number from the nameplate (www.endress.com/deviceviewer) • The Endress+Hauser Operations App: Enter the serial number from the nameplate or scan the 2-D matrix code (QR code) on the nameplate. Standard documentation...
Page 191
Proline Promag W 400 HART Technical data Brief Operating Instructions Brief Operating Instructions for the sensor Measuring device Documentation code Proline Promag W KA01266D Measuring device Documentation code Promag 400 KA01263D Description of device parameters Measuring device Documentation code Promag 400...
Index Proline Promag W 400 HART Index Degree of protection ..... 51, 158 Design Access authorization to parameters Measuring device ......13 Read access .
Page 193
Proline Promag W 400 HART Index Electrical connection Functions Degree of protection ..... . . 51 see Parameter Measuring device ......38 Operating tools Galvanic isolation .
Page 194
Index Proline Promag W 400 HART Length of connecting cable ....24 Line recorder ......120 Nameplate Local display .
Page 195
Proline Promag W 400 HART Index Totalizer 1 to n (Submenu) ....102 Settings Totalizer handling (Submenu) ....119 Adapting the measuring device to the process Web server (Submenu) .
Page 196
Index Proline Promag W 400 HART Simulation ......110 Turning the transmitter housing ....34 Status input .