5-2 Internal Weight Calibration

Turn on and warm up the machine for 15 to 30 minutes before calibration.
Adjust calibration switch to "ON", and the screen will display
TARE
Press
or
Procedure
ENTER
Zero
Calibration
ESC
Weight
Calibration
ESC
Calibration
Switch "OFF"

Zero Calibration

a ) Ensure nothing on the platter; after being stable, press
.......
"
". The zero calibration will be complete 5 seconds later.
b ) To abort zero calibration, just press
Weight Calibration
a ) Place an object, whose weight is known, on the platter, and input the weight value from
front panel. After being stable, press
weight calibration will be complete 5 seconds later.
b ) To abort weight calibration, just press
FM User Manual
ZERO
to select
ENTER
weight mass
.
.
ENTER
Zero calibration complete
Place the
ENTER
and set the
weight
.
ESC
and the screen will display "
ENTER
.
ESC
16
.
Confirmed
ENTER
, and the screen will display
.......
SME300000119
". The