Holding Displayed Values And Displaying Min/Max Values - YOKOGAWA CA700-01 User Manual

Ca700 series pressure calibrator
Table of Contents

Advertisement

6.4
Holding Displayed Values and Displaying Min/
Max Values
Procedure
Holding Displayed Values
1.
When measured values are displayed on the screen, press HOLD on the CA700 front panel.
HOLD appears on the screen, and the measured values are held. You will only be able to use the HOLD
and
2.
To release the hold, press HOLD again.
Min/Max Value Display
1.
When measured values are displayed on the screen, press MIN/MAX on the CA700 front
panel.
The minimum and maximum values are displayed on the screen.
2.
To clear them, press MIN/MAX again.
Explanation
Holding Displayed Values
You can use the hold feature to hold the display. While the display is held, "HOLD" is displayed at the
top of the screen, and you will only be able to use the HOLD and
Min/Max Value Display
The min/max feature displays the minimum and maximum values of the measured pressure, current
and voltage.
This feature is not available during leak pressure measurement.
The following features are disabled while minimum and maximum values are displayed.
• Auto power-off
• Measurement range change
• Relative display (based on the reference value)
• Zero calibration
IM CA700-01EN
keys.
Holding
The minimum and maximum values
keys.
1
2
3
4
5
6
7
Index
6-7

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ca700-02Ca700-03

Table of Contents