5.5

Status Byte Format

Device Status Byte
Status byte format (see the explanation of the <ESC S> command)
bit 7
0 (fixed)
bit 7: Fixed to 0.
bit 6: Fixed to 1.
bit 5: Set to 1 when an error occurs during output.
The information is retained until the status byte is read.
bit 4: Set to 1 when an error occurs in the 24 V loop measurement power supply.
The information is retained until the status byte is read.
bit 3: Set to 1 when a measurement over-range occurs.
In auto-range operation, this bit is set to 1 when the CA700 switches to the higher range.
The information is retained until the status byte is read.
bit 2: Set to 1 when a prohibited operation or command is processed, when the command cannot be
interpreted, or when a parameter is outside the setting range.
The information is retained until the status byte is read.
bit 1: Set to 1 when the output stabilizes after it is changed with the output turned on.
The information is retained until the status byte is read.
bit 0: Set to 1 when measured data is confirmed during measurement.
The information is retained until the status byte is read.
5-18
bit 6
bit 5
1 (fixed)
Output error 24 V power
bit 4
bit 3
Over-range
supply error
bit 2
bit 1
Syntax error Output
change
completion
bit 0
Measurement
complete
IM CA700-01EN