Generating Dc Current And Dc Voltage - YOKOGAWA CA700-01 User Manual

Ca700 series pressure calibrator
Table of Contents

Advertisement

2.3

Generating DC Current and DC Voltage

Procedure
Refer to the Getting Started Guide (IM CA700-02EN), and correctly connect the CA700 to the current
or voltage supply device.
DC current and DC voltage source is valid when the pressure measurement of Display1 is set
to normal pressure measurement (P or EXT P). Follow the procedure in section 2.1 to set the
measurement pressure to normal pressure (P or EXT P).
1.
From the top menu, use the ▲ or ▼ key to select Measure and Source, and press ENTER.
The source/measure screen appears.
2.
When Display2 is showing the measure screen, press SOURCE/MEASURE to switch Display2
to SOURCE.
3.
Press Setting (F5).
Setting categories appear on the function menu.
4.
Press Display2 Setting (F2).
The source function setting screen appears.
5.
Press ENTER.
The measurement item options are displayed.
6.
Use the ▲ or ▼ key to select the source function, and press ENTER.
7.
Press Setting Done(F5).
The screen shown in step 1 appears.
IM CA700-01EN
Display1
Switches to the SOURCE display
(SOURCE mA, SIMULATE mA, SOURCE V)
Display2
Function menu
Set the source function.
Confirms the settings
1
2
3
4
5
6
7