3.5
Setting the Calibration Conditions of the
Pressure Switch
Procedure
1.
In the calibration display, press Select Proc. (F2).
A list of registered calibration procedures is displayed.
Up to 20 sets of procedures can be registered. They are displayed in four windows.
2.
Press
edit. Press ENTER.
Information of the selected calibration procedure is displayed.
3.
Press Edit (F1).
The calibration procedure setting display appears.
3-16
or
to change the page and ▲ or ▼ to select the calibration procedure you want to
IM CA700-01EN