3.2

Select the Calibration Procedure

Procedure
1.
From the top menu, use the ▲ or ▼ key to select Calibrate, and press ENTER.
The calibration display appears.
2.
Press Select Proc.(F2).
A list of registered calibration procedures is displayed.
Up to 20 sets of procedures can be registered. They are displayed in four windows.
3.
Press
edit. Press ENTER.
Information of the selected calibration procedure is displayed.
Explanation
Calibration Procedure
You can register up to twenty sets of calibration processes. A calibration procedure consists of a set of
device information, measurement function and source function.
To calibrate a pressure device, select one of the registered procedures to perform calibration.
IM CA700-01EN
or
to change the page and ▲ or ▼ to select the calibration procedure you want to
Information about the selected calibration
procedure (source function, measurement
function, number of calibration points,
and calibration direction)
1
2
3
4
5
6
7
Index
3-3