Commands - YOKOGAWA CA700-01 User Manual

Ca700 series pressure calibrator
Table of Contents

Advertisement

5.4

Commands

Command Syntax
The command syntax is explained below.
Setting/Control
Command:
Answer:
Queries
Command:
Return:
Mode-specific conditions
Normal:
Adjustment: Can be used in adjustment mode.
[AG command] Sets or queries averaging.
AG
[AP command] Sets or queries the auto power-off feature.
AP
[BGD command] Queries the calibration date.
BGD
[BL command] Sets or queries the on/off state of the backlight.
BL
IM CA700-01EN
Transmission command format
Response data format without response (setting/control)
When an error occurs, the same data as the error message ERRm (m = error
number) that is displayed on the screen is returned.
Transmission command format
Response data format with response (queries)
Can be used for normal measurement/source operation.
Sets or queries averaging.
Command = AGmn<CRLF>→Answer = AGmn<CRLF>
Command = AGm?<CRLF>→Return = AGmn<CRLF>
Parameters
m =
0: GROUP item A
1: GROUP item B
n =
0: OFF
1: ON
Sets or queries the auto power-off feature.
Command = APm<CRLF> → Answer = APm<CRLF>
Command = AP?<CRLF> → Return = APm<CRLF>
Parameters
m =
0: OFF
1: 60 min (default value)
Queries the calibration date.
Command = BGD?<CRLF> → Answer = BGDm,yyyymmdd<CRLF>
Parameters
m = 0: Delivery inspection date
1: Factory calibration date
2: User calibration date
yyyy: Year 4byte, mm: Month 2byte, dd: Day 2byte
Sets or queries the on/off state of the backlight.
Command = BLm<CRLF> → Answer = BLm<CRLF>
Command = BL?<CRLF> → Return = BLm<CRLF>
Parameters
m = 0: Off (default value)
1: On
1
2
3
4
5
6
Normal Adjustment
Yes
No
7
Index
Normal Adjustment
Yes
No
Normal Adjustment
Yes
Yes
Normal Adjustment
Yes
Yes
5-5

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ca700-02Ca700-03

Table of Contents