Advertisement

Quick Links

Products
Solutions
Services
Operating Instructions
Liquiline To Go CYM290

Advertisement

Table of Contents
loading

Summary of Contents for Endress+Hauser Liquiline To Go CYM290

  • Page 1 Products Solutions Services Operating Instructions Liquiline To Go CYM290...
  • Page 3: Basics

    Basics Warranty Defects occurring within 3 years from delivery date shall be remedied free of charge at our plant (carriage and insurance paid by sender). Subject to change. Return of Products Under Warranty Please contact our Service Team before returning a defective device. Ship the cleaned device to the address you have been given.
  • Page 4: Table Of Contents

    Table of Contents Basics ..........................3 Table of Contents .......................4 Package Contents ................6 Documentation ................... 7 Overview ..................... 8 Value-Added Features ....................9 Protective Cover ....................... 10 Hook ..........................10 Display and Keypad ....................11 Start-Up ..................... 12 Inserting the Batteries ................... 12 Connecting a Sensor ....................
  • Page 5 Info and Help Texts ....................42 Error Messages ......................43 pH Product Line ................44 pH Sensors ......................... 44 Endress+Hauser Buffer Solutions (pH) ............45 Ready-to-use quality pH buffer solutions............45 Accessories for pH ....................45 Conductivity Product Line ............... 46 Conductivity Sensors .....................
  • Page 6: Package Contents

    Package Contents Check the shipment for transport damage and completeness. The package of the Liquiline To Go CYM290 includes: Liquiline To Go CYM290 Meter incl. 4 batteries (AA) and premounted quiver Carrying strap Data carrier with detailed user manuals USB cable, 1.5 m...
  • Page 7: Documentation

    Documentation Specific Test Report CD-ROM Complete documentation: • User manuals in different languages • Safety instructions • Certificates • Quickstart guides Safety Instructions In official EU languages and others. • EU Declarations of Conformity CAUTION! These safety instructions are part of the product documentation and must be observed.
  • Page 8: Overview

    Overview The Liquiline To Go CYM290 is a portable multiparameter meter for use with analog or digital sensors. The meter automatically recognizes a connected Memosens sensor and accordingly selects the correspond- ing process variable. By simply replacing the Memosens sensor, the meter can be used for measuring conductivity, pH or oxygen.
  • Page 9: Value-Added Features

    Overview Value-Added Features Memosens The Liquiline To Go CYM290 can communicate with Memosens sensors. These digital sensors are automatically identified and the meter switches to the appropriate measurement method. When a Memosens sensor is connected to the meter, it is indicated by the logo shown on the right.
  • Page 10: Protective Cover

    Overview Protective Cover The front of the meter is protected by a cover, which can be completely flipped over and secured to the back for operation. Hook A fold-out hook on the back allows suspending the meter. This leaves your hands free for the actual measurement.
  • Page 11: Display And Keypad

    Overview Display and Keypad Display and keypad correspond directly via softkeys. Menu icons Menu selection Information Calibration Information Return Start logger Data logger Configuration Softkeys Softkeys Function is shown in the display above the key Arrow keys Selecting / Adjusting entries Confirming an adjustment Switching on / off Switching on / Immediate access of meas.
  • Page 12: Start-Up

    A special lithium-ion battery suited to the battery compartment is available for the Liquiline To Go CYM290. The battery is recharged through the USB port. A battery icon in the display indicates the battery power level: Icon fully filled...
  • Page 13: Connecting A Sensor

    Start-Up Connecting a Sensor The Liquiline To Go CYM290 provides a pH socket acc. to DIN 19262 for analog pH sensors. Alternatively, you can connect a Memosens sensor for pH, conductivity or oxygen measurement. The meter automatically recognizes a connected Memosens sensor and accordingly selects the corresponding process variable.
  • Page 14: Switching On The Meter

    Start-Up Switching On the Meter You can use to switch the meter on: • With , the meter immediately switches to measuring mode. • With , first a self test is performed and then the calibration data and settings are displayed before the meter switches to measur- ing mode.
  • Page 15: Information

    Information Oxy Cond 1) In measuring mode, press the Menu softkey. 2) Select “Information” and confirm by pressing E. 3) Select the desired submenu and confirm by pressing E. The different submenus are described below. Calibration Record Shows the data of the last calibration performed on the currently connected sensor. Sensor Information (Digital Sensors only) Shows the data of the currently connected digital sensor.
  • Page 16: Sensor Network Diagram (Ph And Oxy Only)

    Information Sensor Network Diagram (pH and Oxy only) Provide single-glance information on the following parameters of the connected sensor: • Slope • Zero point (operating point for Memosens ISFET) • Sensocheck (pH) or leakage current (ISFET and Oxy) • Response time •...
  • Page 17: Sensor Monitor

    Information Oxy Cond Sensor Monitor Shows the raw values available fom the connected sensor: mV, temperature, temperature detector, temperature pH, analog resistance pH, digital, glass mV, temperature, glass impedance pH, digital, ISFET mV, leakage current, temperature pH, ORP mV, temperature Cond, digital Resistance, conductance, temperature Sensor current, leakage current, polarization voltage,...
  • Page 18: Device Info

    Information Oxy Cond Device Info Shows the following device information: • Device name • Serial number • Software version • Hardware version • Air pressure • Battery Device Test A device self-test is automatically run in the background at regular intervals. It checks the memory modules listed below.
  • Page 19: Configuration

    Off | On Cal timer Interval On: 00 … 99 days Programmed buffer | Manual | Data entry Cal mode Auto | 1-point | 2-point | 3-point Calibration points Endress+Hauser Buffer set 2.00/4.01/6.98/9.95/11.87 Mettler-Toledo 2.00/4.01/7.00/9.21 Knick CaliMat 2.00/4.00/7.00/9.00/12.00 Ciba 2.06/4.00/7.00/10.00 NIST technical 1.68/4.00/7.00/10.01/12.46...
  • Page 20 Configuration “Configuration” menu selection – part 2 + Time/Date* 24 h | 12 h Time format dd.mm.yyyy | yyyy-mm-dd | dd/mm/yyyy | Date format mm/dd/yyyy Time hh:mm:ss Date Date format as configured + Display* Modern | b/w Appearance Permanent | 60 min | 30 min | 10 min | Lighting 5 min | 1 min | 30 sec Bright | Standard | Dim...
  • Page 21 Configuration Cond Conductivity Configuration 1) In measuring mode, press the Menu softkey. 2) Select “Configuration” and confirm by pressing E. 3) Make the desired adjustments. The following table gives you an overview. Factory settings are shown in bold print. “Configuration” menu selection – part 1 Language Deutsch | English | Español | Italiano | Français | Português...
  • Page 22 Configuration Cond “Configuration” menu selection – part 2 + Calibration* Auto | Entry of solution | Cell constant Cal. mode CLY 11-A | CLY 11-B | CLY 11-C | CLY 11-D | Cal. solution CLY 11-E + Time/Date* 24 h | 12 h Time format dd.mm.yyyy | yyyy-mm-dd | dd/mm/yyyy | Date format...
  • Page 23 Configuration Cond “Configuration” menu selection – part 3 + Data logger* Without Meas.point Without Note Right softkey Logger Start/Stop | Hold value Non-circular | Circular Recording Logger type Shot Interval 1…12:59:59 | 00:02:00 Difference 1st difference On | Off Delta cond 0 …...
  • Page 24 Configuration Oxygen Configuration 1) In measuring mode, press the Menu softkey. 2) Select “Configuration” and confirm by pressing E. 3) Make the desired adjustments. The following table gives you an overview. Factory settings are shown in bold print. “Configuration” menu selection – part 1 Deutsch | English | Español | Italiano | Français | Language Português...
  • Page 25 Configuration “Configuration” menu selection – part 2 + Display* Modern | b/w Appearance Permanent | 60 min | 30 min | 10 min | Lighting 5 min | 1 min | 30 sec Bright | Standard | Dim Brightness + Data logger* Without Meas.point Without...
  • Page 26: Calibration

    (Automatic calibration with specification of the buffer solution used) 1) Select the number of calibration points and the buffer set as shown in the table below and press the Start softkey. Calibration points Auto | 1-point | 2-point | 3-point Endress+Hauser 2.00/4.01/6.98/9.95/11.87 Mettler-Toledo 2.00/4.01/7.00/9.21 Knick CaliMat 2.00/4.00/7.00/9.00/12.00...
  • Page 27: Manual Calibration

    Calibration Manual Calibration (Calibration with manual specification of the number of calibration points and the buffer solution) 1) Select the number of calibration points and press the Start softkey. 2) Adjust the temperature-corrected value (see buffer table) for the 1st/2nd/3rd buffer solution and press Continue (repeat this step for each calibration point).
  • Page 28: Auto Calibration

    Calibration Cond Conductivity Calibration 1) In measuring mode, press the Menu softkey. 2) Select “Configuration” and confirm by pressing E. 3) Select the desired “Calibration mode”. 4) Perform the selected calibration as described on the following pages. Follow the instructions on the display. Auto Calibration (Automatic calibration with specification of the calibration solution used) 5) Select the calibration solution:...
  • Page 29: Entry Of Solution" Calibration

    Calibration Cond ”Entry of Solution” Calibration (Calibration by entering the conductivity with display of the cell constant) 1) Press the Start softkey. 2) Immerse the sensor in the solution. 3) Enter the temperature-corrected conductivity value and press E. 4) Finally, you can Apply these values or Cancel the calibration. Cell Constant Calibration (Calibration by entering the cell constant with display of the conductivity) 1) Press the Start softkey.
  • Page 30: Calibration In Air

    Calibration Oxygen Calibration 1) In measuring mode, press the Menu softkey. 2) Select “Calibration” and confirm by pressing E. 3) Select desired “Calibration mode” and confirm by pressing E. 4) Select “Membrane module replacement” if you wish to save a change of membrane or electrolyte in the connected sensor.
  • Page 31: Zero Calibration

    Calibration Zero Calibration (Zero calibration with oxygen-free medium, e.g., nitrogen 5.0) 1) Place sensor in oxygen-free medium and wait for a stable measured value. 2) Press Start softkey. Calibration will be performed. 3) Finally, you can Apply these values or Cancel the calibration. Data Entry Calibration (Calibration by entering known sensor values) 1) Press Start softkey.
  • Page 32: Measuring

    Measuring Oxy Cond Once you have completed all preparations, you can start with the actual measure- ment. 1) Connect the desired sensor to the meter. Some sensors require a special prepara- tion. Please proceed according to the operating instructions for the sensor. 2) Switch the device on using the key.
  • Page 33: Data Logger

    Data Logger Oxy Cond The Data Logger The meter provides a data logger. Prior to use, it must be configured and then activated. You can choose from the following logger types: • Shot (manual logging by pressing the Save value softkey) •...
  • Page 34: Operating Modes Of The Data Logger (Logger Type)

    Data Logger Oxy Cond Operating Modes of the Data Logger (Logger Type) Shot In this mode, a measured value is recorded when the Save value softkey is pressed. In the measuring mode ( ), it is always possible to hold a value and then save it. Interval (time-controlled) In the “Interval”...
  • Page 35 Data Logger Oxy Cond Interval and difference (combined) When the delta range related to the last DIFF entry is exceeded, a new entry is creat- ed (example: entry A) and the delta range is displaced upwards or downwards by the delta value.
  • Page 36: Configuring The Data Logger

    Data logger Oxy Cond Configuring the Data Logger Prerequisite: Data logger is stopped. The “Data logger” menu shows the number of occupied entries as well as the number of free entries. Configuration can also be done in the “Configuration” menu under “Data logger”.
  • Page 37 Data Logger Oxy Cond Configuring the data logger (default in bold print) Without Meas.point Without Note Logger Start/Stop | Hold value Right softkey Non-circular Recording Circular Logger type Shot Interval 00:00:01…12:59:59 | 00:02:00 Interval Difference 1st difference pH 0.0…16.0 | pH 1.0 Delta pH 0 …...
  • Page 38: Starting/Stopping The Data Logger

    Data Logger Oxy Cond Starting/Stopping the Data Logger With the data logger activated, automatic switch-off is disabled. Every time the meter has been switched off, the data logger must be restarted. Depending on the assignment of the right softkey (see “Configuring the data logger”), you can start/stop the data logger as follows: Right softkey 1) Press right softkey Start logger / Stop logger.
  • Page 39 Data Logger Oxy Cond Example 1: Viewing the logger data Data Logger Primary mea- Measured sured value temperature Date and time Meas.point Note Sensor Analog Scroll using arrow keys Graphic Return Example 2: Curve characteristic Data Logger Select entry using arrows.
  • Page 41: Error And Status Messages

    Error and Status Messages Oxy Cond Error and status messages appear as plain text on the display. By pressing E and Help, more detailed help texts can be displayed. Information on the sensor condi- tion is indicated by the “Sensoface” icon (friendly, neutral, sad) possibly accompanied by an info text.
  • Page 42: Info And Help Texts

    Error and Status Messages Oxy Cond “Sensoface” Messages The “Sensoface” icon provides information on the sensor condition: Sensoface Meaning Sensor is okay Calibrate the sensor soon Calibrate or replace the sensor Info and Help Texts When an error or status message appears on the screen, proceed as follows to view the corresponding info or help text: 1) Press E.
  • Page 43: Error Messages

    Error and Status Messages Oxy Cond Error Messages Error Message Remedy Replace the batteries Replace the batteries. blinks ERR 1 Primary variable range Check whether the measurement conditions ERR 2 ORP range correspond to the adjusted measuring range. ERR 3 Temperature range ERR 4 Zero point...
  • Page 44: Ph Product Line

    pH Product Line pH Sensors Product Name Sensor type (Link to Product Configurator) Orbisint CPS11D Digital pH sensor Orbisint CPS12D Digital ORP sensor Memosens CPS16D Combined digital pH/ORP sensor Memosens CPS31D Digital pH sensor Ceraliquid CPS41D Digital pH sensor Ceraliquid CPS42D Digital ORP sensor Ceragel CPS71D Digital ORP sensor...
  • Page 45: Endress+Hauser Buffer Solutions (Ph)

    Ready-to-use quality pH buffer solutions Quality buffers from Endress+Hauser - CPY20 Solutions which are traced by a DAkkS-accredited Endress+Hauser buffer laboratory (DkkS = German Accreditation Body) to a primary reference material of the PTB and to standard reference material of the National Institute of Standards and Technol- ogy (NIST) in accordance with DIN 19266 are used as secondary reference buffer solutions.
  • Page 46: Conductivity Product Line

    Conductivity Product Line Cond Conductivity Sensors Product Name Sensor type (Link to Product Configurator) Condumax CLS15D Conductivity sensor Condumax CLS16D Conductivity sensor Condumax CLS21D Conductivity sensor The Product Configurator can be accessed at: www.endress.com/<product name> Memosens sensors have a cable coupling, which allows convenient replacement of sensors while the cable remains connected to the meter.
  • Page 47: Conductivity Calibration Solutions Cly11

    Conductivity Product Line Cond Conductivity Calibration Solutions CLY11 CLY11-A, 74 μS/cm (reference temp. 25°C (77 °F)), 500ml (16,9 fl.oz) CLY11-B, 149.6 μS/cm (reference temp. 25°C (77 °F)), 500ml (16,9 fl.oz) CLY11-C, 1.406 mS/cm (reference temp. 25°C (77 °F)), 500ml (16.9 fl.oz) CLY11-D, 12.64 mS/cm (reference temp.
  • Page 48: Oxygen Product Line

    Oxygen Product Line Oxygen Sensor Product Name (Link to Product Configurator) Oxymax COS22D digital oxygen sensor The Product Configurator can be accessed at: www.endress.com/<product name> Accessories for Oxygen Item COS22D maintenance kit Sensor cable CYK20-AAB1C2 1.5 meters Please visit our website for more information on our product range: www.endress.com.
  • Page 49: Specifications

    Specifications Oxy Cond Connections 2x socket, 4 mm dia., for separate temp. detector 1x M8 socket, 4 pins, for Memosens lab cable 1x micro USB-B 1x pH socket, to DIN 19262 Air pressure measurement 700 … 1100 hPa User interface Straightforward menu navigation with graphic icons and detailed operating instructions in plain text Languages...
  • Page 50 Specifications Oxy Cond Communication USB 2.0 Profile HID, driverless installation Usage Data exchange Diagnostics functions Sensor data Manufacturer, sensor type, serial number, wear, operating time (Memosens only) Calibration data Calibration date; pH/Oxy: zero, slope; Cond: Cell constant Device self-test Automatic memory test (FLASH, EEPROM, RAM) Device data Device type, software version, hardware version Data retention...
  • Page 51 Specifications Analog pH/mV input pH socket, DIN 19 262 (13/4 mm) pH range -2 … 16 Decimal places * 2 or 3 Input resistance 1 x 10 Ω (0 … 35 °C) Input current 1 x 10 (at RT, doubles every 10 K) Measuring cycle Approx.
  • Page 52 Programmed buffers Calibration with automatic buffer recognition Manual Manual calibration with entry of individual buffer values Data entry Data entry of zero and slope Buffer sets * -01- Endress+Hauser 2.00/4.01/6.98/9.95/11.87 -02- Mettler-Toledo 2.00/4.01/7.00/9.21 -03- Knick CaliMat 2.00/4.00/7.00/9.00/12.00 -04- Ciba (94) 2.06/4.00/7.00/10.00 -05- NIST technical 1.68/4.00/7.00/10.01/12.46...
  • Page 53 Specifications Cond Conductivity input M8 socket, 4 pins, for Memosens lab cable Measuring Ranges CLS15D k = 0.01 : 0-20 µS/cm k = 0.1 : 0-200 µS/cm CLS16D k = 0.1 : 0.04 µS/cm - 500 µS/cm CLS21D k = 1 : 10.0 µS/cm - 20.0 mS/cm Permissible cell constant 0.005 …...
  • Page 54 Specifications Memosens input, oxygen M8 socket, 4 pins, for Memosens lab cable Display ranges Saturation 0.000 … 200.0 % Concentration 000 μg/l … 20.00 mg/l Partial pressure 0.0 … 1000 mbar Temperature meas. range -20 … 150 °C Sensor standardization Automatic calibration in air (100 % RH) Zero calibration Storage...
  • Page 55: Index

    Index AA batteries 12 Accessories for conductivity 47 Accessories for oxygen 48 Accessories for pH 45 Arrow keys 11 Auto, conductivity calibration 28 Automatic calibration (conductivity) 28 Automatic calibration (pH) 26 Basic interval (limit value logger type) 35 Battery capacity 12 Battery compartment 12 Battery icon 12 Battery life, increasing 36...
  • Page 56 Index Certificates 7 Charge level of batteries 12 CIP (sensor information) 15 Clearing the logger 38 Compensation (conductivity configuration) 21 Conductivity calibration 28 Conductivity calibration solutions CLY11 47 Conductivity configuration 21 Conductivity sensors, product line 46 Configuration (Cond) 21 Configuration (Oxy) 24 Configuration (pH) 19 Configuring the data logger 36 Connecting a sensor 13...
  • Page 57 Index Digital sensors, pH 44 Display 11 Display icons 14 Displaying the logger data 38 Disposal 3 Documentation 7 Electrolyte, Oxy calibration 30 Electrolyte vessel, CPY7 45 Entering a tag number (Oxy) 30 Entry of solution, conductivity calibration 29 ERROR (error codes) 43 Error messages, overview 43 EU Declarations of Conformity 7 Event interval (limit value logger type) 35...
  • Page 58 Index Limit value (logger type) 35 Lithium-ion battery (start-up) 12 Logger type “Difference” 34 Logger type “Interval” 34 Logger type “Interval and difference” 35 Logger type “Limit value” 35 Logger type “Shot” 34 Manual calibration (pH) 27 Measuring 32 Measuring point (Oxy calibration) 30 Measuring point (TAG) 17 Membrane 30 Membrane module replacement 30...
  • Page 59 Index Package contents 6 pH buffer solutions 45 pH calibration 26 pH configuration 19 pH sensors, product line 44 Point of measurement, saving (Oxy calibration) 30 Ports 13 Power-on 14 Pressure correction (Oxy configuration) 24 Pre-trigger (limit value logger type) 35 Product features 8 Product line 44 Product presentation 8...
  • Page 60 Index Sensor network diagram 16 Sensors for conductivity, product line 46 Sensors for oxygen, product line 48 Sensors for pH, product line 44 Sensor without temperature detector 32 Serial number of sensor (sensor information) 15 Setup, conductivity 21 Setup, oxygen 24 Setup, pH 19 Shot (logger type) 34 SIP (sensor information) 15...
  • Page 61 Index USB port (batttery) 12 USB port, micro 13 Value-added features 9 Viewing the logger data 38 Warranty 3 Zero calibration (Oxy) 31...
  • Page 64 Endress+Hauser Conducta GmbH+Co. KG Dieselstraße 24 70839 Gerlingen - Germany Phone: +49 7156 209 790 Fax: +49 7156 28 158 088995 TA-209.7MU-EHE01 20150413 Software version: 1.x 71265350...

Table of Contents