Endress+Hauser Smartec S CLD134 Operating Instructions Manual

Conductivity measuring system
Hide thumbs Also See for Smartec S CLD134:
Table of Contents

Advertisement

Quick Links

Operating Instructions
Smartec S CLD134
Conductivity Measuring System
R
Standard Number 74-03
BA401C/07/en/07.07
71059820
Software version 1.13

Advertisement

Table of Contents
loading

Summary of Contents for Endress+Hauser Smartec S CLD134

  • Page 1 Operating Instructions Smartec S CLD134 Conductivity Measuring System Standard Number 74-03 BA401C/07/en/07.07 71059820 Software version 1.13...
  • Page 2 Spare parts that can be delivered and a system overview are listed on these pages. Æ Technical data → Page 78 Dimensions → Page 78 ff. Process conditions, weight, material Æ Index → Page 84 ff. The index helps you to find information and important terms easily and quickly. Endress+Hauser...
  • Page 3: Table Of Contents

    3.3.3 CLD 134 compact version ... . . 17 Maintenance of Smartec S CLD134 ... . 59 Installation instructions ..... 19 7.1.1 Dismantling Smartec S CLD134 .
  • Page 4 Smartec S CLD134 Technical Data....76 10.1 Input ........76 10.2 Output .
  • Page 5: Safety Instructions

    Safety instructions Safety instructions Designated use Smartec S CLD134 is a field-tested and reliable transmitter used to determine the conductivity of liquid media. It is particularly suitable for use in the foodstuffs industry. Any other use than the one described here compromises the safety of persons and the entire measuring system and is, therefore, not permitted.
  • Page 6: Notes On Safety Conventions And Symbols

    Safety instructions Smartec S CLD134 Notes on safety conventions and symbols Safety symbols Warning! This symbol alerts you to hazards. They can cause serious damage to the instrument or to persons if ignored. Caution! " This symbol alerts you to possible faults which could arise from incorrect operation. They could cause damage to the instrument if ignored.
  • Page 7: Identification

    Smartec S CLD134 Identification Identification Device designation 2.1.1 Nameplate Compare the order code on the nameplate (on the Smartec) with the product structure (see below) and check that it agrees with your order. You can identify the instrument variant by the order code on the nameplate. Under "Codes", you can find the release code for the software upgrade "MRS".
  • Page 8: Product Structure Smartec S Cld134

    Identification Smartec S CLD134 2.1.2 Product structure Smartec S CLD134 Housing Transmitter only (without sensor) Compact version Separate transmitter, cable length 5 m / 16.41 ft Separate transmitter, cable length 10 m / 32.81 ft Separate transmitter, cable length 20 m / 65.62 ft...
  • Page 9: Basic Version And Function Extensions

    Biological reactivity according to USP <87>, <88> class VI Scope of delivery The scope of delivery of the compact version inlcudes: • Smartec S CLD134 compact measuring system with integrated sensor • Terminal strip set • Operating Instructions BA401C/07/en • Versions with HART communication only: Operating Instructions Field communication with HART, BA212C/07/en •...
  • Page 10: Certificates And Approvals

    Identification Smartec S CLD134 Certificates and approvals Declaration of conformity The product meets the legal requirements of the harmonized European standards. The manufacturer confirms compliance with the standards by affixing the 4 symbol. All materials in contact with medium are listet at FDA.
  • Page 11: Installation

    • If you have not yet installed the sensor at the measuring point, perform an Airset and install the sensor (see the Technical Information of the sensor). • Connect the sensor to the Smartec S CLD134 as described in the chapter "Electrical connection". • Connect the transmitter as described in the chapter "Electrical connection".
  • Page 12: Incoming Acceptance, Transport, Storage

    Installation Smartec S CLD134 Incoming acceptance, transport, storage • Make sure the packaging is undamaged! Inform the supplier about damage to the packaging. Keep the damaged packaging until the matter has been settled. • Make sure the contents are undamaged! Inform the supplier about damage to the delivery contents.
  • Page 13 Smartec S CLD134 Installation Wall distance The sensor’s distance from the pipe wall affects the measuring accuracy (see Fig. 5). In narrow installation conditions, the ion flow in the medium is affected by the pipe walls. This effect is compensated by the so-called installation factor.
  • Page 14: Cld134 Separate Version

    Installation Smartec S CLD134 3.3.2 CLD134 separate version 142 / 5.59 Ø 7 / 0.28 225 / 8.86 175 / 6.89 mm /inch a0005632 Fig. 6: CLD134 wall mounting with mounting plate 109 / 4.29 Ø max. 60 / Ø max. 2.36 170 / 6.69...
  • Page 15 Smartec S CLD134 Installation a0005429 Fig. 8: Dimensions CLS54 (long version) Conductivity sensors for the separate transmitter CLS54 conductivity sensors with various process connections covering all common installation conditions are available for the separate version. a0004949 Fig. 9: Process connections CLS54 (short version) NEUMO BioControl D50 Varivent N DN 40 ...
  • Page 16 Installation Smartec S CLD134 a0005436 Fig. 10: Process connections CLS54 (long version) Dairy pipe fitting DIN 11851, DN 50 (union nut is included) SMS 2" (union nut is included) Clamp ISO 2852, 2" Aseptic- fitting DIN 11864-1 form A, for pipe according to DIN 11850, DN 50...
  • Page 17: Cld 134 Compact Version

    Smartec S CLD134 Installation 3.3.3 CLD 134 compact version a0005500 Fig. 11: Dimensions of CLD134 compact version depending on ordered process connection Process connections Various process connections covering all common installation conditions are available for the compact version. The compact version is installed at the measuring point with the required process connection.
  • Page 18 Installation Smartec S CLD134 a0005502 Fig. 13: Process connections compact version (long) Dairy pipe fitting DIN 11851 DN 50 (union nut is included) SMS 2" (union nut is included) Clamp ISO 2852, 2" Aseptic fitting DIN 11864-1 form A, for pipe according to DIN 11850, DN 50...
  • Page 19: Installation Instructions

    Smartec S CLD134 Installation Installation instructions 3.4.1 Mounting CLD134 separate version Wall mounting For wall mounting, attach the mounting plate to the wall by drilling holes as required. Anchors and screws are to be provided by the operator. 142 / 5.59 Ø...
  • Page 20 Installation Smartec S CLD134 Post mounting A mounting kit for installing the housing on horizontal or vertical posts or pipes (max. Ø 60 mm / Ø 2.36") is available as an accessory (see chapter "Accessories"). a0004902 Fig. 15: Mounting kit for installing the CLD134 separate version on posts Note! For use in hygienic sensitive areas, shorten the threads as much as possible.
  • Page 21: Mounting Cld134 Compact Version Or Cls54 Sensor For Separate Version

    Install the compact version or the CLS54 sensor directly on the pipe or vessel socket via the process connection (depending on ordered version). When installing the Smartec S CLD134 or the sensor, make sure that the flow opening of the sensor is oriented in the flow direction of the medium. An orientation arrow on the sensor facilitates orientation.
  • Page 22: Wiring

    • Ensure that there is no voltage at the power cable before beginning the connection work. 4.1.1 Electrical connection of transmitter Proceed as follows to connect the Smartec S CLD134: 1. Loosen the 4 Phillips screws on the housing cover and remove the cover.
  • Page 23 Smartec S CLD134 Wiring a0005439 Fig. 19: Terminal assignments of cable glands on Smartec S CLD134 Separate version Compact version Cable gland for analog output, binary input Cable gland for analog output, digital input Cable gland for alarm contact Cable gland for alarm contact...
  • Page 24 15 V 10-50 V – – 10-50 V a0004895 Fig. 21: Electrical connection of Smartec S CLD134 Signal output 1 conductivity Conductivity sensor Signal output 2 temperature Temperature sensor Auxiliary power output Alarm (contact position: no current) Binary input 2 (MRS1+2)
  • Page 25 Smartec S CLD134 Wiring Connection compartment sticker 131082-4A Sensor 31 + 32 - 33 + Temp. (opt.) 34 - 85 + +15V 10mA 86 - 93 + Bin 2 94 - 81 + Bin 1 82 - – Mains Hilfsenergie –...
  • Page 26: Alarm Contact

    Wiring Smartec S CLD134 Structure and termination of measuring cable Semiconductor (84) (83) screen Screen Screen 16 15 84 (15) (16) GN WH YE GN (11) WH (12) CLK5 (unused) YE (13) CLK5 a0005640-en a0004906 Fig. 24: Structure of CLK5 measuring cable Fig.
  • Page 27: Post-Connection Check

    Smartec S CLD134 Wiring Post-connection check After wiring up the electrical connection, carry out the following checks: Device status and specifications Remarks Are the transmitter or the cable externally damaged? Visual inspection Electrical connection Remarks Are the installed cables strain-relieved? No loops and cross-overs in the cable run? Are the signal cables correctly connected acc.
  • Page 28: Operation

    (PLC). Note! For operation via HART or PROFIBUS PA/DP, read the corresponding chapters in the additional operating instructions: • PROFIBUS PA/DP, field communication with Smartec S CLD134, BA213C/07/en ® • HART , field communication with Smartec S CLD134, BA212C/07/en The following chapters describe local operation via operating keys.
  • Page 29: Operating Elements

    SMARTEC S ALARM – a0004899 Fig. 29: Operating elements of Smartec S CLD134 Liquid crystal display showing measured values and configuration data 4 operating keys for calibration and instrument configuration Field for user labeling LED indicator for alarm function 5.2.3...
  • Page 30 Operation Smartec S CLD134 PLUS key and MINUS key In setup mode, the PLUS and MINUS keys have the following functions: • Selection of function groups Note! To select function groups in the order given in the chapter "Instrument configuration", use the MINUS key.
  • Page 31: Local Operation

    Smartec S CLD134 Operation Local operation 5.3.1 Operating concept Operating modes Measuring mode: standard mode of operation, displaying current measured values Setup mode Calibration mode: access to all execution of calibration routine configuration settings Code The function groups are selected with the PLUS or MINUS key.
  • Page 32 Operation Smartec S CLD134 Menu structure The configuration and calibration functions are arranged in a menu structure by function groups. The function groups are selected in the setup mode with the PLUS and MINUS keys. The ENTER key is used to move from one function to the next within a function group.
  • Page 33: Commissioning

    C111 top to the bottom and from the left to the right. C07-CLD132xx-13-06-00-xx-005.eps Fig. 33: Function coding For a detailed description of the function groups available on the Smartec S CLD134 see the chapter "Instrument configuration". Endress+Hauser...
  • Page 34 Commissioning Smartec S CLD134 Factory settings When the instrument is switched on for the first time, the factory settings are in effect. The following table provides an overview of all major settings. Please refer to the description of the individual functions in the chapter "Instrument configuration"...
  • Page 35: Quick Setup

    Smartec S CLD134 Commissioning Quick setup After switching the transmitter on, configure the major functions required for accurate measurement. The following section gives you an example for a basic configuration. Input Selection or range Display (factory setting bold) 1. Press the ENTER key.
  • Page 36 Commissioning Smartec S CLD134 Input Selection or range Display (factory setting bold) 14. In A6, press the ENTER key to confirm the factory setting. 1.000 If your wall distance is smaller than 15 mm / 0.59", 0.10 ... 1 ... 5.00 refer to the chapters 3.3.1 and 6.4.14 for information...
  • Page 37 Smartec S CLD134 Commissioning Input Selection or range Display (factory setting bold) 26. In O2, select the linear characteristic. lin = linear (1) Confirm your selection by pressing ENTER. sim = simulation (2) Sel.Type 27. In O211, select the current range for your output, 4-20 4 ...
  • Page 38: Instrument Configuration

    Commissioning Smartec S CLD134 Instrument configuration The following sections give a detailed description of all Smartec S CLD134 functions. 6.4.1 Setup 1 (conductivity, concentration) In the SETUP 1 function group, you can change the operating mode and the sensor settings.
  • Page 39: Setup 2 (Temperature)

    Smartec S CLD134 Commissioning 6.4.2 Setup 2 (temperature) The temperature compensation only needs to be performed in the conductivity mode (selection in field A1). The temperature coefficient specifies the change in conductivity per degree of temperature change. It depends on the chemical composition of the medium and the temperature itself.
  • Page 40 Commissioning Smartec S CLD134 k( ) C07-CLD132xx-05-06-00-xx-011.eps Fig. 36: Determination of temperature coefficient Required data Calculated α values Use the following formula to calculate the α values for the temperatures occurring in your process: (T) - (T ) a = 100% ¹...
  • Page 41: Current Outputs

    Smartec S CLD134 Commissioning Coding Field Selection or range Display Info (factory settings bold) ° 25.0 Enter process 25 °C Only if B1 = fixed. temperature -10.0 ... 150.0 °C This value can only be specified in °C. ProcTemp. Display temperature Display and entry of real °...
  • Page 42: Alarm

    Commissioning Smartec S CLD134 Coding Field Selection or range Display Info (factory settings bold) Enter the measured value corresponding to Cond: 2000 mS/cm 20 mA value: the maximum current value (20 mA) at the mS/cm 2000 Conc: 99.99 % O213...
  • Page 43: Check

    Smartec S CLD134 Commissioning Coding Field Selection or range Display Info (factory settings bold) Select the errors that are to trigger an alarm signal. The errors are selected via the error number. Select error Please refer to the table in chapter 9.2 "System 1 ...
  • Page 44: Relay Configuration

    (Error no.: E152) 6.4.6 Relay configuration For Smartec S CLD134 equipped with remote parameter set switching (measuring range switching), there are three options for configuring the relay (selection in field R1): • Alarm The relay closes the contact 41/42 (voltage-free, safe state) if an alarm condition according to chapter 9.2 occurs and if the setting in the “Alarm contact”...
  • Page 45 Smartec S CLD134 Commissioning C07-CLD132xx-05-06-00-xx-008.eps Fig. 38: Relation of switch-on and switch-off points and pickup and dropout delays Switch-on point > switch-off point: Max. function Switch-on point Switch-on point < switch-off point: Min. function Switch-off point Contact ON Contact OFF...
  • Page 46: Temperature Compensation With Table

    Commissioning Smartec S CLD134 Coding Field Selection or range Display Info (factory settings bold) Enter dropout delay 0 ... 2000 s Off Delay auto auto This selection can only be made if "limit" has Select simulation manual been selected in R1.
  • Page 47: Concentration Measurement

    Smartec S CLD134 Commissioning Coding Field Selection or range Display Info (factory settings bold) The temperature values must have a minimum distance of 1 K. °C Enter temperature 0.0 °C Factory setting for temperature value of value value -10.0 ... 150.0 °C pairs in table: Temp.val.
  • Page 48 Commissioning Smartec S CLD134 κ κ C07-CLD132xx-05-06-00-xx-012.eps Fig. 39: Measured data for variable process temperatures (example) κ Conductivity Measuring point Concentration Measuring range Temperature C07-CLD132xx-05-06-00-xx-015.epsB Fig. 40: Measured data for a constant process temperature (example) κ Conductivity Constant temperature Concentration...
  • Page 49 Smartec S CLD134 Commissioning • Enter the values in the order of increasing concentration (see the following example). mS/cm °C (°F) 60 (140) 90 (194) 60 (140) 90 (194) 60 (140) 90 (194) Coding Field Selection or range Display Info (factory settings bold) Settings for concentration measurement.
  • Page 50: Service

    Commissioning Smartec S CLD134 Coding Field Selection or range Display Info (factory settings bold) mS/cm Enter uncompensated 0.0 mS/cm conductivity 0.0 ... 9999 mS/cm conduct. Enter concentration 0.00 % value for K6 0.00 ... 99.99 % concentr. °C Enter temperature 0.0 °C...
  • Page 51 Smartec S CLD134 Commissioning Coding Field Selection or range Display Info (factory settings bold) Manual hold Man.HOLD Enter hold dwell 10 s period 0 ... 999 s Cont.Time Enter SW upgrade Entering an incorrect code returns you to the release code of measurement menu.
  • Page 52: E+H Service

    Commissioning Smartec S CLD134 6.4.10 E+H Service Coding Field Selection or range Display Info (factory settings bold) Function group E+H E+H service settings. SERVICE E+H SERV Contr = controller (1) Contr Trans = transmitter (2) Select module MainB = mainboard (3)
  • Page 53: Determining The Temperature Coefficient

    Smartec S CLD134 Commissioning 6.4.12 Determining the temperature coefficient Determining the temperature coefficient by the following method is only possible for instruments equipped with remote parameter set switching (see "Product structure"). Standard instruments (basic versions) can be retrofitted with remote parameter set switching (see chapter "Accessories").
  • Page 54 Commissioning Smartec S CLD134 Assignment of binary inputs The transmitter has 2 binary inputs. They can be defined in field M1 as follows: Assignment of field M1 Assignment of binary inputs M1 = 0 MRS not active. The binary input 1 can be used for external hold.
  • Page 55 Smartec S CLD134 Commissioning Coding Field Selection or range Display Info (factory settings bold) cond. cond = conductivity The operating mode can be individually Select operating mode conc = concentration defined for each parameter set. Oper.Mode NaOH NaOH, H2SO4, H3PO4, HNO3 Select medium Only available if M4 = conc.
  • Page 56: Calibration

    Commissioning Smartec S CLD134 6.4.14 Calibration To access the "Calibration" function group, press the CAL key. This function group is used to calibrate the transmitter. Two different types of calibration are possible: • Calibration by measurement in a calibration solution of a known conductivity.
  • Page 57 Smartec S CLD134 Commissioning Coding Field Selection or range Display Info (factory settings bold) If C113 = E xxx, then only no or new. Store calibration C114 C114 If new, return to C. results? If yes/no, return to “Measurement”. Store...
  • Page 58: Communication Interfaces

    Commissioning Smartec S CLD134 Coding Field Selection or range Display Info (factory settings bold) Airs = Airset (1) Calibration with InstF Cellc = cell constant (2) C1 (3) sensor adaptation for InstF = installation Sensor calibration with compensation of wall...
  • Page 59: Maintenance

    • For your own safety, use only original spare parts. Original parts will guarantee functionality, accuracy and reliability after repairs. Note! Please contact your Endress+Hauser representative if you have any questions. You can also send your queries to the Endress+Hauser Service Organisation via the Internet: www.endress.com Maintenance of Smartec S CLD134 7.1.1...
  • Page 60: Replacement Of Central Module

    – Monitoring functions – Interface parameters Dismantle the instrument as described in the chapter "Dismantling Smartec S CLD134". Refer to the part number of the central module to determine whether the new module has the same part number as the old one.
  • Page 61: Maintenance Of Measuring System

    Smartec S CLD134 Maintenance Maintenance of measuring system 7.2.1 Cleaning conductivity sensors Inductive sensors are less sensitive to soiling than conventional conductive sensors since there is no galvanic contact with the medium. However, dirt may collect in the flow opening (making it narrower), which changes the cell constant.
  • Page 62: Instrument Check By Medium Simulation

    Maintenance Smartec S CLD134 7.2.3 Instrument check by medium simulation The inductive sensor cannot be simulated. However, the overall system comprising the CLD134 and inductive sensor can be checked using equivalent resistances. Note the cell constant ( k = 6.3 cm for CLS54).
  • Page 63: Checking Line Extension And Junction Box

    Smartec S CLD134 Maintenance 7.2.4 Checking line extension and junction box • Use the methods described in chapters "Checking inductive conductivity sensors" and "Instrument check by medium simulation" to perform a quick functional check from the conductivity sensor to the measuring instrument via an extension.
  • Page 64: Accessories

    Accessories Smartec S CLD134 Accessories Sensors • Indumax H CLS54 Inductive conductivity sensor with fast response time and hygienic design; with integrated temperature sensor. Order according to product structure, see Technical Information TI400C/07/en. Extension cable • Extension cable CLK5 for inductive conductivity sensors, for extension via the VBM junction box, sold by the meter;...
  • Page 65: Post Mounting Kit

    Smartec S CLD134 Accessories Post mounting kit • Mounting kit for installation of Smartec S CLD132/CLD134 on horizontal or vertical pipes and posts (max. Ø 60 mm (2.36")), material stainless steel 1.4301; order no.: 50062121 a0004902 Fig. 42: Mounting kit for installing CLD132/CLD134 separate version on posts or pipes (base plate belongs to the...
  • Page 66: Troubleshooting

    Optoscope, see chapter "Service equipment not match hardware (central module) Optoscope"). E002 4. If problem persists, return instrument to your local Endress+Hauser sales agency for repair or replace instrument. Download error Download must not access locked functions (e.g. E003 temperature table in basic version).
  • Page 67: Process-Specific Errors

    Smartec S CLD134 Troubleshooting Error no. Display Tests and / or measures Alarm contact Error current Factory User Factory User E063 Below current output range 1 Check measured value and current output assignment (function group O). E064 Current output range 1 exceeded...
  • Page 68 Troubleshooting Smartec S CLD134 Error Possible cause Tests and / or remedial measures Equipment, spare parts, personnel Short circuit / moisture in sensor Check sensor. See chapter "Checking inductive conductivity sensors". Short circuit in cable or junction box Check cable and junction box.
  • Page 69 Smartec S CLD134 Troubleshooting Error Possible cause Tests and / or remedial measures Equipment, spare parts, personnel Relay configured for alarm Activate limit contactor. See field R1. Pickup delay setting too long Shorten pickup delay. See field R4. Limit contact does not work "Hold"...
  • Page 70: Instrument-Specific Errors

    Troubleshooting Smartec S CLD134 Error Possible cause Tests and / or remedial measures Equipment, spare parts, personnel No central PA/DP module Verify by looking at nameplate: Upgrade to LSCP module, PA = -xxxxxxPEx and = -xxxxxxPFx see chapter "Spare parts".
  • Page 71 Smartec S CLD134 Troubleshooting Error Possible cause Tests and / or remedial measures Equipment, spare parts, personnel Ribbon cable or transmitter module not Reinsert transmitter module, use additional Refer to exploded view in chapter "Spare Display shows measured properly installed fastening screw M3 if necessary.
  • Page 72: Spare Parts

    Troubleshooting Smartec S CLD134 Spare parts Spare parts are to be ordered from your sales centre responsible. Specify the order numbers listed in the chapter "Spare parts kits". To be on the safe side, you should always specify the following data with your spare part orders: •...
  • Page 73: Exploded View

    Pos. 100 /101 box during assembly a0005652-en The exploded view drawing shows all components and spare parts of Smartec S CLD134. Use the position numbers to find the spare parts designation and their order number in the following section. Endress+Hauser...
  • Page 74: Spare Part Kits

    Troubleshooting Smartec S CLD134 9.5.2 Spare part kits Item Kit designation Name Function/content Order number Housing bottom, separate Bottom assembly 51501574 Housing bottom, compact Bottom assembly 51501576 Post mounting kit 1 pair of post mounting parts 50062121 Housing cover Cover with accessories...
  • Page 75: Return

    Smartec S CLD134 Troubleshooting Return If the transmitter has to be repaired, please return it cleaned to the sales centre responsible. Please use the original packaging, if possible. Please enclose the completed "Declaration of contamination" (copy the second last page of these Operating Instructions) with the packaging and also the shipping documents.
  • Page 76: Technical Data

    Technical Data Smartec S CLD134 Technical Data 10.1 Input Measured variables Conductivity Concentration Temperature Measuring range Conductivity: recommended range: 100 μS/cm ... 2000 mS/cm (uncompensated) Concentration – NaOH: 0 ... 15 % – HNO 0 ... 25 % – H 0 ...
  • Page 77: Power Supply

    –10 ... +70 °C / 14 ... 158 °F (separate version) –10 ... +55 °C / 14 ... 131 °F (compact version) (see Fig. 41 "Permissible temperature ranges of Smartec S CLD134") Storage temperature –25 ... +70 °C / -13 ... 158 °F Electromagnetic compatibility Interference emission and interference resistance acc.
  • Page 78: Mechanical Construction

    Technical Data Smartec S CLD134 10.6 Mechanical construction Design, dimensions Separate transmitter with mounting L x W x D: 225 x 142 x 109 mm (8.86 x 5.59 plate x 4.29 ") Compact transmitter MV5, CS1, L x W x D: 225 x 142 x 109 mm (8.86 x 5.59 AA5, SMS: x 10.04 ")
  • Page 79: Chemical Durability Of Cls54 Sensor

    Smartec S CLD134 Technical Data Permissible temperature ranges of Smartec S CLD134 [°F] [° C] –10 [° C] –10 [°F] Medium temperature a0005499-en Fig. 43: Permissible temperature ranges of Smartec S CLD134 CLS54 sensor with separate version Compact version Short term for sterilisation (<60 min)
  • Page 80: Appendix

    Appendix Smartec S CLD134 Appendix Operating matrix Entry of Entry of correct Display of calculated Entry of a value of Calibration calibration temperature conductivity value of installation factor calibration solution (if B1 = fixed) calibration solution InstF 2.10 %/K current meas. value C133 25.0 C...
  • Page 81 Smartec S CLD134 Appendix Store Display of calibration results calibration status ; no; new o.k.; E--- C136 C135 Display of Store calibration status calibration results o.k.; ; no; new E--- C125 C126 Entry of installation factor Entry of measured value damping 01 ...
  • Page 82 Appendix Smartec S CLD134 Selection of contact Function group Selection of Selection of contact Pickup delay Dropout delay function switch-on point RELAY switch-off point setting setting (with software option only) 2000 mS/cm; 99.99 % Alarm 2000 mS/cm; 99.99 % Limit;...
  • Page 83 Smartec S CLD134 Appendix Selection of Switch simulation simulation on or off (only if R1 = limit) (only if R6 = manual) auto manual Entry of Output temperature coefficient a table status o.k. (y value) ; no 2.10 %/K 0.00 ... 20.00 %/K...
  • Page 84: Index

    Declaration of conformity ......10 Smartec S CLD134 ......59 Dismantling .
  • Page 85 Smartec S CLD134 PCS alarm ........43 Performance characteristics .
  • Page 86 Smartec S CLD134 Endress+Hauser...
  • Page 87 Erklärung zur Kontamination und Reinigung Please reference the Return Authorization Number (RA#), obtained from Endress+Hauser, on all paperwork and mark the RA# clearly on the outside of the box. If this procedure is not followed, it may result in the refusal of the package at our facility.