Endress+Hauser Prosonic S FMU90 Brief Operating Instructions

Flow measurement backwater and dirt detection totalizers and counters
Hide thumbs Also See for Prosonic S FMU90:

Advertisement

KA01068F/00/EN/02.16
71308093
Products
Brief Operating Instructions
Prosonic S FMU90
Flow Measurement
Backwater and Dirt Detection
Totalizers and Counters
These Instructions are Brief Operating Instructions; they are not
a substitute for the Operating Instructions pertaining to the
device.
Detailed information about the device can be found in the
Operating Instructions and the other documentation:
Available for all device versions via:
– Internet:
www.endress.com/deviceviewer
– Smart phone/tablet: Endress+Hauser Operations App
Solutions
Services

Advertisement

Table of Contents
loading

Summary of Contents for Endress+Hauser Prosonic S FMU90

  • Page 1 Operating Instructions pertaining to the device. Detailed information about the device can be found in the Operating Instructions and the other documentation: Available for all device versions via: – Internet: www.endress.com/deviceviewer – Smart phone/tablet: Endress+Hauser Operations App...
  • Page 2: Serial Number

    Prosonic S FMU90 Flow Measurement PROFIBUS Order code: XXXXX-XXXXXX Ser. no.: XXXXXXXXXXXX Ext. ord. cd.: XXX.XXXX.XX Serial number www.endress.com/deviceviewer Endress+Hauser Operations App A0023555 Endress+Hauser...
  • Page 3: Table Of Contents

    Prosonic S FMU90 Flow Measurement PROFIBUS Table of contents Table of contents 1 Safety Instructions ............. 4 1.1 Designated use .
  • Page 4: Safety Instructions

    Prosonic S FMU90 Flow Measurement PROFIBUS Safety Instructions Designated use The Prosonic S FMU90 is a transmitter for the ultrasonic sensors FDU90, FDU91, FDU91F, FDU92, FDU93 and FDU95 . The sensors of the class FDU8x can be connected as well.
  • Page 5: Safety Symbols

    Prosonic S FMU90 Flow Measurement PROFIBUS Safety Instructions Safety symbols Symbol Meaning DANGER! DANGER This symbol alerts you to a dangerous situation. Failure to avoid this situation will result in serious or fatal injury. A0011189-DE WARNING! WARNING This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in serious or fatal injury.
  • Page 6: Mounting

    Mounting Prosonic S FMU90 Flow Measurement PROFIBUS Mounting Incoming acceptance, storage 2.1.1 Incoming acceptance Check the packing and contents for any signs of damage. Check the shipment, make sure nothing is missing and that the scope of supply matches your order.
  • Page 7 Further information: Operating Instructions for Prosonic S FMU90, document number BA00293F. You will find this document in the Download Area of the Endress+Hauser Internet site. 2.2.2 Installation conditions of the DIN-rail housing • The DIN-rail housing must be mounted outside hazardous areas in a cabinet.
  • Page 8: Mounting The Remote Display And Operating Module

    The dimensions are described in this document: Technical Information for Prosonic S FMU90, document number TI00397F. You will find this document in the Download Area of the Endress+Hauser Internet site. Mounting the remote display and operating module 2.3.1 Scope of delivery...
  • Page 9 Further information: Operating Instructions for Prosonic S FMU90, document number BA00293F. You will find this document in the Download Area of the Endress+Hauser Internet site. 2.3.4 Adaption plate If an opening of 138 x 138 mm (5.43 x 5.43 in) and the remote display of the Prosonic FMU860/861/862 are already present, you can use the adaption plate (Order Code: 52027441).
  • Page 10: Mounting Of The Sensors

    Mounting Prosonic S FMU90 Flow Measurement PROFIBUS L00-FMU90xxx-06-00-00-xx-006 Remote display of the FMU90 with adaption plate Opening of the remote display of the FMU860/861/862 Mounting of the sensors Information on the mounting of the sensors can be found in the following documents: •...
  • Page 11: Wiring

    Prosonic S FMU90 Flow Measurement PROFIBUS Wiring Wiring WARNING Risk of electric shock! Voltage at the terminals. ‣ Switch off the supply voltage before connecting the device. ‣ In a wet environment, do not open the cover if voltage is present.
  • Page 12: Terminal Compartment Of The Din-Rail Housing

    Wiring Prosonic S FMU90 Flow Measurement PROFIBUS 3.1.1 Cable entries of the field housing The following openings for cable entries are prestamped on the bottom of the housing: M20x1.5 (10 openings), M16x1.5 (5 openings), M25x1.5 (1 opening). The required number and types of cable entries depend on the application at hand.
  • Page 13 Prosonic S FMU90 Flow Measurement PROFIBUS Wiring 3.2.2 Several instruments mounted side by side L00-FMU90xxx-04-00-00-xx-012 • The cables can be inserted into the housing from above or from below. • The pictures show the smallest housing version but are valid for the larger versions as well.
  • Page 14: Terminal Assignment

    Wiring Prosonic S FMU90 Flow Measurement PROFIBUS Terminal assignment Pluggable spring-force terminals for connection of the cables are supplied in the terminal compartment. Rigid conductors or flexible conductors with cable sleeve can directly be inserted and are contacted automatically. Feature...
  • Page 15 Prosonic S FMU90 Flow Measurement PROFIBUS Wiring Relay DigIn 0/4...20mA Sync Service Fuse Display FDU- Sensor HART 0/4…20mA FDU- FDU- Sensor Sensor Relay L00-FMU90xxx-04-00-00-xx-001 Terminals of the Prosonic S (the terminals depicted in grey are not present in every instrument version)
  • Page 16 Wiring Prosonic S FMU90 Flow Measurement PROFIBUS Terminals Meaning Terminal area Remarks Auxiliary energy • L (for AC version) Depending on instrument version: • L+ (for DC version) • 90 to 253 V • N (for AC version) • 10.5 to 32 V •...
  • Page 17 Prosonic S FMU90 Flow Measurement PROFIBUS Wiring Terminals Meaning Terminal area Remarks External switch inputs 71, 72, 73 External switch input 1 0: < 8 V or 72 and 73 interconnected 1: > 16 V or 72 and 73 not interconnected...
  • Page 18: Sensor Connection

    (e.g. PLC) are described in this document: Operating Instructions "PROFIBUS DP/PA - Guidelines for planning and commissioning", document number BA00034S. • Further information: Operating Instructions for Prosonic S FMU90, document number BA00293F. Sensor connection Information on the mounting of the sensors can be found in the following documents: •...
  • Page 19: Connection Of The Sensor Heater (For Fdu90/Fdu91)

    Prosonic S FMU90 Flow Measurement PROFIBUS Wiring Connection of the sensor heater (for FDU90/FDU91) The FDU90 and FDU91 sensors are available in a version with heater. For more information and details, see the Technical Information TI00396F that is delivered with the sensors.
  • Page 20 Wiring Prosonic S FMU90 Flow Measurement PROFIBUS 3.5.2 Connection in the DIN-rail housing The supply voltage must be provided in the cabinet, e.g. by a terminal on the DIN-rail: FMU90 Sync Service Fuse Display HART 0/4…20mA FDU- Sensor Relay FDU90...
  • Page 21: Connection Of External Switches (For Fmu90-********B***)

    The maximum short-circuit current at 24 V is 20 mA. Connection of a temperature sensor The Prosonic S FMU90 transmitter has an optional input for an external temperature probe (in the product structure: feature 90 "Additional input", option B). The following probes can be connected: •...
  • Page 22 Wiring Prosonic S FMU90 Flow Measurement PROFIBUS 3.7.1 FMT131 (Endress+Hauser), (connectable to FMU90-********B***) FMU90 83 84 FMU90 FMU90 83 84 BK YE YEGN YEGN YEGN FMT131-J FMT131-J FMT131-R L00-FMU90xxx-04-00-00-xx-019 Non-Ex area (FMT131-R) black Ex area (FMT131-J) with grounding in the FMU90...
  • Page 23 Prosonic S FMU90 Flow Measurement PROFIBUS Wiring 3.7.2 Pt100 (connectable to FMU90-********B***) 84 85 FMU90 Pt100 Pt100 L00-FMU90xxx-04-00-00-xx-020 Pt100 with 3-wire-connection Pt100 with 4-wire-connection (one connector remains unused) A Pt100 with 2-wire-connection may not be used due to its insufficient measuring accuracy.
  • Page 24: Synchronization Line

    Wiring Prosonic S FMU90 Flow Measurement PROFIBUS Synchronization line • If wiring several Prosonic S (FMU90/FMU95) which are mounted in a common cabinet and if the sensor cables FMU90 FMU90 FMU90 run in parallel, the synchronization terminals (39 and 40) must be interconnected.
  • Page 25: Connection Of The Separate Display And Operating Module

    Prosonic S FMU90 Flow Measurement PROFIBUS Wiring Connection of the separate display and operating module Sync Service Fuse Display HART 0/4…20mA FDU- Sensor Relay L00-FMU90xxx-04-00-00-xx-005 Connection of the display plug with the cable (3 m (9.8 ft)) For the version of the Prosonic S with a separate display for panel mounting, a pre-assembled connecting cable (3 m (9.8 ft)) is supplied.
  • Page 26 Wiring Prosonic S FMU90 Flow Measurement PROFIBUS Example Sync Service Fuse Display HART 0/4…20mA FDU- Sensor Relay GNYE FDU91F/93/95/96 90 … 253 VAC (FDU83/84/85/86) 10,5 … 32 VDC L00-FMU90xxx-04-00-00-xx-007 The wire is already connected on delivery Endress+Hauser...
  • Page 27 Prosonic S FMU90 Flow Measurement PROFIBUS Wiring 3.10.2 Potential equalization for the DIN-rail housing If the DIN-rail housing is used, the potential equalization must be connected in the cabinet, e.g. at a metallic DIN rail: Sync Service Fuse Display HART 0/4…20mA...
  • Page 28: 3.11 Post-Connection Check

    Wiring Prosonic S FMU90 Flow Measurement PROFIBUS 3.11 Post-connection check WARNING Risk of electric shock! Voltage at the terminals. ‣ Switch off the supply voltage before you check the terminal assignment. ‣ In a wet environment, do not open the cover if voltage is present.
  • Page 29: Operation

    "FieldCare" – Via PROFIBUS DP with the PROFIcard, PROFIboard or PROFIusb – Acyclic data exchange, SLOT index tables Further information: Operating Instructions for Prosonic S FMU90, document number BA00293F. Operation via the display and operating module 4.2.1 Display and operating elements...
  • Page 30 Operation Prosonic S FMU90 Flow Measurement PROFIBUS Display symbols Symbol Meaning Operating mode of the instrument User User parameters can be edited. Service parameters are locked. Diagnosis The service interface is connected. Service User and service parameters can be edited.
  • Page 31 Prosonic S FMU90 Flow Measurement PROFIBUS Operation Keys (softkey operation) The function of the keys depends on the current position within the operating menu (softkey functionality). The key functions are indicated by softkey symbols in the bottom line of the display.
  • Page 32 Operation Prosonic S FMU90 Flow Measurement PROFIBUS General key combinations The following key combinations do not depend on the menu position: Key combination Meaning Escape • While editing a parameter: Exit the editing mode without accepting the changes. • Within the navigation: Move upwards to the previous layer of the menu.
  • Page 33 :dome ceiling  medium property:liquid process cond:standard liq. Identification of the parameter sets: Submenu Number of the associated input or output Number of the parameter set within the submenus Further information: Operating Instructions for Prosonic S FMU90, document number BA00293F. Endress+Hauser...
  • Page 34: Commissioning

    Commissioning Prosonic S FMU90 Flow Measurement PROFIBUS Commissioning Setting the device address 5.1.1 Selecting the device address • Every PROFIBUS device must be given an address. If the address is not set correctly, the device will not be recognised by the process control system.
  • Page 35: Bus Termination

    Prosonic S FMU90 Flow Measurement PROFIBUS Commissioning Hardware addressing comes into operation when DIP switch 8 is in the position "HW (OFF)". In this case the address is determined by the position of DIP-switches 1 to 7 according to the following table: Switch No.
  • Page 36: Loading The Device Database And Type Files (Gsd)

    PLC. Additional bitmap files are required in order to represent the device by an icon in the network design software. The device database and bitmap files are needed for the commissioning of a PROFIBUS DP network. Further information: Operating Instructions for Prosonic S FMU90, document number BA00293F, Chapter "Commissioning" First setup...
  • Page 37: Basic Setup

    Prosonic S FMU90 Flow Measurement PROFIBUS Commissioning Basic setup After the first setup the main screen appears (flow or level). However, the displayed values do not yet correspond to the real flows or level before you have performed the basic setup. To do so, enter the main menu by pressing "Menu" (right key). Select the "flow"...
  • Page 38: Overview Basic Setup For A Flow Measurement

    Commissioning Prosonic S FMU90 Flow Measurement PROFIBUS The "level" submenu is used for the calibration of level measurement (1 or 2 channels) Step Parameter Set Parameter Remarks Main menu → level → Level (LVL)1 Select the "level" submenu. In the following submenu →...
  • Page 39 Prosonic S FMU90 Flow Measurement PROFIBUS Commissioning Step Parameter Set Parameter Remarks → Beta Specify the value of the parameter β, (only available for "type" = "formula") → Gamma Specify the value of the parameter γ, (only available for "type" = "formula") →...
  • Page 40 Commissioning Prosonic S FMU90 Flow Measurement PROFIBUS Step Parameter Set Parameter Remarks → Flow N state → Level Displays the currently measured level (N = 1 or 2) → Distance Displays the currently measured distance between the reference point of the sensor and the liquid surface. Check the value: •...
  • Page 41 Prosonic S FMU90 Flow Measurement PROFIBUS Commissioning 5.6.1 Flow N linearization The selected linearization type determines which parameters are present. Only the parameters "type" and "flow unit" are always present. The "linearization" parameter set is used to calculate the flow from the measured level. The Prosonic S provides the following linearization types: •...
  • Page 42 Commissioning Prosonic S FMU90 Flow Measurement PROFIBUS Parameter Selection/Remarks "edit" This parameter is used to enter or to view the linearization table. You have got the following options: Selection • Read The table editor appears. An existing table can be viewed but not changed.
  • Page 43 Prosonic S FMU90 Flow Measurement PROFIBUS Commissioning Blocking distance Parameter Remarks "blocking distance" Indicates the blocking distance of the respective sensor. The blocking distance is measured from the reference point of the respective sensor. For more information and details see the Technical Information TI00396F.
  • Page 44: Overview Basic Setup For Level Measurement

    Commissioning Prosonic S FMU90 Flow Measurement PROFIBUS Overview basic setup for level measurement The following table gives an overview of the basic setup for level measurement. Detailed information on the parameters can be found in the Operating Instructions BA00293F. Step...
  • Page 45 Prosonic S FMU90 Flow Measurement PROFIBUS Commissioning Step Parameter set Parameter Remarks → linearization → Type Select type of linearization (Details → "type") → Mode Specify, to which value the measurement refers: "level" or "ullage" → Customer unit Specify the unit for the linearized value;...
  • Page 46 Commissioning Prosonic S FMU90 Flow Measurement PROFIBUS Step Parameter set Parameter Remarks → LVL 1(2) state → Level 1(2) Indicates the currently measured level. → Act. distance Indicates the currently measured distance between the reference point of the sensor and the product surface.
  • Page 47: Application Parameters

    Prosonic S FMU90 Flow Measurement PROFIBUS Commissioning 5.7.1 Application parameters Parameter Selection/Remarks "tank shape" Use this parameter to specify the tank shape of your application: Selection: • dome ceiling (→ A) • horizontal cyl. (→ B) • bypass (→ C) •...
  • Page 48 Commissioning Prosonic S FMU90 Flow Measurement PROFIBUS "process conditions" for the following situations Example filter settings standard liquid for all fluid applications The filters and output damping are set which do not fit in any of the to average values.
  • Page 49 Prosonic S FMU90 Flow Measurement PROFIBUS Commissioning "process conditions" for the following situations Example filter settings solid dusty Dusty bulk solids The averaging filters are set to detect even relatively weak signals. conveyor belt Bulk solids with rapid level The averaging filters are set to small change values.
  • Page 50 Commissioning Prosonic S FMU90 Flow Measurement PROFIBUS 5.7.2 Empty and full calibration Parameter Remarks "empty calibration" Use this parameter to specify the empty distance E, i.e. the distance between the reference point of the sensor and the minimum level (zero point).
  • Page 51 Prosonic S FMU90 Flow Measurement PROFIBUS Commissioning 5.7.3 Linearization Parameter Selection/Remarks "type" Number and type of the parameters in this set depend on the selected linearization type. Only the parameters "type" and "mode" are always present. Select the linearization type in this parameter.
  • Page 52 Commissioning Prosonic S FMU90 Flow Measurement PROFIBUS Parameter Selection/Remarks "type" In these linearization types the measured level is convertet to the volume in a horizontal cylinder or a spherical tank. Selection • Horizontal cylinder • Sphere The following additional parameters have to be specified: •...
  • Page 53 Prosonic S FMU90 Flow Measurement PROFIBUS Commissioning Parameter Selection/Remarks "type" In this linearization mode the measured value is calculated from a linearization table. The table may consist of up to 32 pairs of values (level - volume). The table must be monotonically increasing or decreasing.
  • Page 54 Commissioning Prosonic S FMU90 Flow Measurement PROFIBUS Parameter Selection/Remarks The table editor Level Level Value Value 0,0000 0,0000 0,0000 0,0000 0,0000 0,0000 0,0000 0,0000 0,0000 0,0000 0,0000 0,0000 … 0,0000 … 0,0000 0,0000 0,0000 : go to next row : navigate within the table : go to previous row : (for "Level"...
  • Page 55 Prosonic S FMU90 Flow Measurement PROFIBUS Commissioning 5.7.4 Interference echo suppression: Basic principles Parameter Remarks "check value" The "check value" and "distance mapping" parameters are used to configure the interference "distance mapping" echo suppression of the Prosonic S. The following picture shows the operating principle of the interference echo suppression:...
  • Page 56 71308093 www.addresses.endress.com...