Fluke Endurance 1RL Operating Instructions Manual page 52

2-color infrared thermometer
Table of Contents

Advertisement

Programming Guide
Description
Peak hold time (4)
IP Portaddress
Wide Power
Narrow power
Video relative reticle diameter
Video relative reticle X-position
Video relative reticle Y-position
Slope
Set target temperature
Target Temperature 2-color
Terminator resistor
TCP/IP time out interval
Temperature units (scale)
Poll/burst mode
Target temp: 1-color wide
Web server ON/OFF
Burst string contents (3)
Multidrop address
Low temperature limit
Deadband (7)
Restore factory defaults
Transmissivity
High temperature limit
Sensor initialization
LASER / LED / Video ON / OFF
Sensor model type
0 - 20 mA / 4-20 mA analog output
Sensor firmware revision no.
Sensor analog part revision no.
Setpoint / Relay function
Trigger
Identify unit
Sensor serial number
Attenuation to activate relay
Attenuation for failsafe
52
info@Raytek-Direct.ca
Char
Format
P
B
(2)
(1)
(1)
P
n.n - nnn.n
PORT
n - nnnnn
Q
n.nnnnnnn
R
n.nnnnnnn
RC
n.n – nn.nn
RX
n.n – nn.nn
RY
n.n – nn.nn
S
n.nnn
STT
n.n – nnnn.n
T
n.n - nnnn.n
TR
n
TTI
n - nnn
U
X
V
X
W
n.n - nnnn.n
WS
n
X$
XA
nnn
XB
n.n - nnnn.n
XD
nn
XF
XG
n.nn
XH
n.n – nnnn.n
XI
n
XL
n
XM
X
XO
n
XR
Xn
XRA
Xn
XS
n.n – nnnn.n
XT
N
XU
Varies
XV
nnnnnnnn
V
Y
nn
Z
nn
Rev. E1 Sep 2015
S
Legal Values
(1)
0.0 – 300.0 sec (300 s = ∞)
1 - 65535
0.850 – 1.150
0.0 – 9999.0 (5)
0 = OFF, 1 = ON
0 = ∞, 1 – 240 sec
C or F
B = Burst , P = Polled
(5)
0 = OFF, 1 = ON
000 to 032
0.0–9999.0 (5)
01 – 55 in °C / 01 – 99 in °F
0.10 – 1.10
0.0–9999.0 (5)
0 = flag reset, 1 = flag set
0 = OFF, 1 = ON
L = Low Temp., H = Hi Temp
0 = 0 - 20 mA, 4 = 4 - 20 mA
e.g. 1.02.11
e.g. 1.02.01
0.0 to 3200.0°C / 5792.0°F (8)
0 = inactive, 1 = active
e.g. E1RL-F2-V-0-0
e.g. 31712345 (8 digits)
0 to 95% energy
0 to 99% energy reduction
ENDURANCE 1R
Factory Default
0.0
6363
1.000
Set at factory calibration
0 = OFF
0
non-US: C
P = Polled
0 = OFF
000
Set at factory calibration
02
1.00
Set at factory calibration
1
0 = OFF
Set at factory calibration
4
Set at factory calibration
Set at factory calibration
0.0
0
Set at factory calibration
Set at factory calibration
95%
95%
1.800.868.7495

Advertisement

Table of Contents
loading

This manual is also suitable for:

Endurance 1rh

Table of Contents