Installation
Multiple Endurance sensors in a RS485 Multidrop Network
Wiring
For an installation of two or more Endurance sensors in a RS485 network (2-wire, half duplex), each
Endurance sensor needs it's specific RS485 network address (1 - 32), preset via the Endurance control
panel (user interface) or alternatively via a standard terminal program (operating system dependent).
Once all the units are addressed, wire up the units in the 2-wire multidrop manner, keeping all A & B
signals to be common and connect these to the both signal lines of the selected USB/RS485 converter.

5.2.6 Addressing the Endurance sensor in a RS485 Multidrop Network

If you are installing two or more sensors in a multi-drop configuration, please be aware of the
following:
• Each sensor must have a unique address greater zero (1 - 32).
• Each sensor must be set to the same baud rate (default is 38.4 kBaud).
• Once all the units are addressed, wire up the units in the 2-wire multidrop manner, keeping all
A & B to be common.
• Now you can run own written communication software to access the Endurance sensor or use
an individual terminal program to issue commands and receive the response from the sensor.
28
info@Raytek-Direct.ca
Figure 18: USB/RS485 Converter
Rev. E1 Sep 2015