Safety Precautions - AGFA Drystar 5500 Reference Manual

X-ray
Hide thumbs Also See for Drystar 5500:
Table of Contents

Advertisement

Safety precautions

The Drystar 5500 must only be operated according to its specifications and its
intended use. Any operation not corresponding to the specifications or intended use
may result in hazards, which in turn may lead to serious injuries or fatal accidents
(for example electric shocks). AGFA positively will not assume any liability in these
cases.
All images created using any image technology can show artifacts which could be
mixed up with diagnostic relevant information. If there is any doubt that the
diagnostic information could not be absolutely true, additional investigations must
be performed to get a clear diagnostic.
When operating or maintaining the Drystar 5500, always observe the
following safety guidelines:
Have electrical or mechanical defects repaired by skilled personnel only!
Do not override or disconnect the integrated safety features.
Ventilation openings may not be covered.
Always switch off the Drystar 5500 and disconnect the power cord from
the outlet before carrying out any maintenance work.
Film jam removal or Cleaning the printer thermal head can be done without
switching the power off. Nevertheless, care should be taken and the following
instructions should be respected:
Always take into account the markings provided on the inside and outside of
the printer. A brief overview of these markings and their meaning is given
below.
Safety warning, indicating that the Drystar 5500 manuals should be
consulted before making any connections to other equipment. The
use of accessory equipment not complying with the equivalent safety
requirements of this printer may lead to a reduced level of safety of
the resulting system. Consideration relating to the choice of
accessory equipment shall include:
• Use of the accessory equipment in the patient vicinity,
• Evidence that the safety certification of the accessory equipment
In addition all configurations must comply with the medical
electrical systems standard IEC 601-1-2. The party that makes the
connections acts as system configurator and is responsible for
complying with the systems standard.
If required contact your local service organization.
2900H EN 20071108
has been performed in accordance with the appropriate IEC 601-1
and IEC 601-1-2 harmonized national standard.
D
5500/5503 R
RYSTAR
EFERENCE MANUAL
Introducing the Drystar 5500
13

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Drystar 5503

Table of Contents