Charts for mammography QC (optional)

Quality Control for
Mammography applications
Drystar 5500/5503 : Determination of Operating Levels
Imager Type: __________
Film Type: ____________
Step 1: Print QC Test images on five consecutive days. Record the optical densities
measurements in the tables below. After five days, average the values to determine
the reference levels for each of the parameters.
Month
Date
Initials
Base + Fog
Average of 5 values = calculated reference "Base + Fog" level
Low Density
Average of 5 values = calculated reference "Low Density" level
Mid Density
Average of 5 values = calculated reference "Mid Density" level
High Density
Average of 5 values = calculated reference "High Density" level
Step 2: Copy the calculated reference levels to Charts 2A/B ('Daily Density Control Chart')
2900H EN 20071108
Serial #: _________________
Emulsion #: ______________
Densitometer : _________________ (default selection)
Day 1
Day 2
D
RYSTAR
Date ____________________
Input Tray: _______________
Day 3
Day 4
5500/5503 R
EFERENCE MANUAL
Chart 1
Day 5
Quality Control Charts
247