AGFA Drystar 5500 Reference Manual page 156

X-ray
Hide thumbs Also See for Drystar 5500:
Table of Contents

Advertisement

D
5500/5503 R
RYSTAR
After the image is printed, the system will display all measured optical density
4
values:
0.19
0.50
1.30
Default densitometer
e.g. Macbeth TR924
The displayed values (that have to be monitored) represent the following steps on the
test film:
Operating Level
Base + Fog
Low density
Mid density
High density
If the mid density value does not meet or exceeds the recommended
values, the cause must be found and the problem solved before any further
clinical films can be printed.
Refer to
page
local Agfa service organization.
Record the density levels on Chart 1 ('Drystar 5500/5503: Determination of
5
Operating Levels'). Refer to
page 247.
Press the Confirm key to return to the main menu.
6
Repeat steps
7
the Drystar 5500 Chart 1.
Calculate the average value of the densities from the five images. These values
8
represent operating levels or aim values, for each density.
156
Advanced operation (Key-operator mode)
EFERENCE MANUAL
the density value of the
Base + Fog step
the density value of the
low density step
the density value of the
mid density step
the density value of the
high density step
'Maintaining image quality and resolving Image quality problems'
220
and
'Preventive maintenance schedule'
'Charts for mammography QC (optional)'
1
through
6
once a day for five consecutive days, as indicated on
Value (Macbeth units)
(according to NEMA
standards XR 23-2006)
0.22 ± 0.03
0.52 ± 0.07
1.35 ± 0.15
2.35 ± 0.15
on page 186, or call your
2.23
3.60
1.73
on
on
2900H EN 20071108