Remote Timeout; Lost Can Partner - Agilent Technologies 1260Infinity II Bio-inert FractionCollector User Manual

Infinitylab lc series
Table of Contents

Advertisement

Remote Timeout

Error ID: 0070
A not-ready condition is still present on the remote input. When an analysis is
started, the system expects all not-ready conditions (for example, a not-ready
condition during detector balance) to switch to run conditions within one
minute of starting the analysis. If a not-ready condition is still present on the
remote line after one minute the error message is generated.
Probable cause
1
2
3

Lost CAN Partner

Error ID: 0071
During an analysis, the internal synchronization or communication between
one or more of the modules in the system has failed.
The system processors continually monitor the system configuration. If one or
more of the modules is no longer recognized as being connected to the system,
the error message is generated.
Probable cause
1
2
3
1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual
Not-ready condition in one of the
instruments connected to the remote line.
Defective remote cable.
Defective components in the instrument
showing the not-ready condition.
CAN cable disconnected.
Defective CAN cable.
Defective main board in another module.
Error Information
General Error Messages
Suggested actions
Ensure the instrument showing the not-ready
condition is installed correctly, and is set up
correctly for analysis.
Exchange the remote cable.
Check the instrument for defects (refer to the
instrument's documentation).
Suggested actions
• Ensure all the CAN cables are connected
correctly.
• Ensure all CAN cables are installed
correctly.
Exchange the CAN cable.
Switch off the system. Restart the system, and
determine which module or modules are not
recognized by the system.
6
95