Replace the Inlet/Waste Tubings (Analytical and Bio-inert)

For bio-inert modules use bio-inert parts only!
When
• When contaminated, worn out or visibly damaged
• Typically once every year
Parts required
#
1
1
1
Preparations
• Position the transport unit of the fraction collector in the Home position.
• Remove all installed trays from the tray base.
• Position the transport unit of the fraction collector in the Change Parts position.
• Turn OFF the instrument.
• Remove the rear end of the fraction collector's waste tubing from the waste container, unscrew
Personal injury
WA R N I N G
Risk of personal injury caused by the needle arm movement.
1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual

Replace the Inlet/Waste Tubings (Analytical and Bio-inert)

p/n
G1364-68602
G1364-68601
G1364-68603
the front end of the fraction collector's inlet tubing from the flow cell of the detector.
Keep fingers away from the needle area.
Description
Tubing Kit 1 – 5 mL/min, 0.25 mm ID
Standard tubing kit for Analytical and Bio-inert Fraction Collector
Tubing Kit 1 mL/min, 0.15 mm ID (OPTIONAL)
Tubing Kit 4 – 8 mL/min, 0.5 mm ID (OPTIONAL)
7
Maintenance
117