Installing a new Automation Interface

If a defective automation interface has been replaced:
1 Disconnect all connections to the Automation Interface
2 Remove the two lock screws holding the Automation Interface (see
3 Withdraw the Automation Interface from the high performance
4 Install the new automation interface as described
1 Turn on the Agilent 1260 Infinity LC system including the high
2 Start the ChemStation SW. Switch to the Diagnosis view and select
3 Type the serial number of the new automation interface and reset the high
Agilent 1260 Infinity Sample Capacity Extension User Manual
Figure
11)
autosampler and replace the lock screws in position in the Automation
Interface.
Automation Interface"
performance autosampler. The autosampler will abort the initialization and
the status LED will turn red.
Maintenance > High Performance ALS > New Automation Interface.
performance autosampler.
on page 19.
Installation
in"Installing the