HP 1650B Programming Reference Manual page 269

Logic analyzers
Table of Contents

Advertisement

RUNTil
Command Syntax:
where:.
Examples:
HP 1650B/HP 16518
Programming Reference
The RUNTY (run until) command defines stop criteria based on the time
between the X and 0 markers when the trace mode is in repetitive. When
key is pressed or the STOP command is sent. Run until the time between
X and 0 marker options are:
Less Than (LT) a specified time value
l
Greater Than (GT) a specified time value
l
In the range (INRange) between two time values
l
Out of the range (OUTRange) between two time values
l
End points for the INRange and OUTRange should be at least 10 11s apart
since this is the minimum time at which data is sampled.
This command affects the timing analyzer only, and has no relation to the
RUNTil commands in the SLISt and COMPare subsystems.
The RUNTil query returns the current stop criteria.
: : = { O F F 1 LT, <value > 1 GT, < v a l u e > 1 INRange < v a l u e > , <value z 1
<value > , <value > }
:: = real number
OUTPUT
O U T P U T XXX;":MACHINEl:TWAVEFORM:RUNTIL
RUNTil
command/query
G T , 800.OE-6"