Retrieving The Alarm Types - GE Optica Operator's Manual

General eastern dew point analyzer
Hide thumbs Also See for Optica:
Table of Contents

Advertisement

January 2006
Retrieving the Alarm
Types
G-4
This command enables the user to retrieve different alarm types that
are supported by the Optica, and the order in which they are
referenced.
Command:
http://3.112.160.36:28005/OpticaAPI.xml?GetAlarmTypes
Function Name: GetAlarmTypes
Channel: 0
Response:
<?xml version="1.0" encoding="UTF-8" standalone="yes" ?>
<OpticaAPIReturn Function="GetAlarmTypes">
<NumAlarmTypes>6</NumAlarmTypes>
<AlarmType>SetPoint</AlarmType>
<AlarmType>Inner Band</AlarmType>
<AlarmType>Outer Band</AlarmType>
<AlarmType>Control</AlarmType>
<AlarmType>Service</AlarmType>
<AlarmType>Pacer</AlarmType>
<retval>GEIAPI_SUCCESS</retval>
</OpticaAPIReturn>
This Response indicates that 6 alarm types are supported. They are
listed in order with the reference being 0.