Imaging Assembly - Konica Minolta DRYPRO SIGMA Service Manual

Medical printer
Hide thumbs Also See for DRYPRO SIGMA:
Table of Contents

Advertisement

SERVICE MANUAL

Imaging Assembly

Prerequisites:
T20 Torx driver
Tools:
Removal
[1] Remove 6 connectors from the MCS board:
• 4 across the bottom of the board (J19, J20, J21, J22)
• 2 on the right side of the board (J15, J17)
[2] Remove 2 screws on the right of the MCS board bracket.
2-38
• Close the film cartridge. (Touch the Pause button on the display screen.)
• De-energize the laser imager.
• Disconnect the power cord.
• Remove:
Top cover
Right cover
Left cover
• To support the imaging assembly during the procedure, move the front and
rear shipping brackets into their shipping location (see Installation
Instructions,
"Moving the Shipping
bracket.
Connectors
Brackets"). Use at least 2 screws per
Screws
9G9002 – 2011-08-15