Konica Minolta DRYPRO SIGMA Service Manual page 143

Medical printer
Hide thumbs Also See for DRYPRO SIGMA:
Table of Contents

Advertisement

Condition Code: 27-607 – Optics Maximum Laser Power Outside Valid Range
Web Portal Message
Optics: Calibration
Failed
Description
System Impact
User Action
Show at DS
DICOM Printer
Status
Unique Symb. ID
Logged for RMS
Condition Code: 27-608 – Invalid Laser Dynamic Range
Web Portal Message
None
Description
System Impact
User Action
FE Action
Show at DS
DICOM Printer
Status
Unique Symb. ID
9G9002 – 2011-08-15
Log Message
Optics calibration
detected Optics
Maximum Power
outside valid range
This error occurs when the maximum laser power measured by the optics
calibration and diagnostics is outside the predefined valid range.
The error clears when a calibration is requested or power is cycled.
The imager can only print jobs that were in process when the error occurred.
The spooler can still accept jobs, but the jobs will not print until the error is
resolved. All jobs that were in process when the error occurred will reprint
completely when the error is resolved. [SI-1]
1. Shut down the laser imager, and then restart the imager.
2. If the error persists, call for service.
True
FAILURE
ERR_OPTICS_MAX_LASER_POWER
[TriggerValues=x][TriggerRange=y,z][TriggerNotes=Laser Power: value, limits]
Log Message
Optics calibration
detected laser dynamic
range approaching
predefined limits
This notification occurs when the laser dynamic range has been adjusted and
its value is within 10% of the upper or lower limit of the adjustment range.
None. The system will function normally. [SI-0]
None
None. This is a log notification only.
False
NORMAL
ERR_OPTICS_LASER_DYNAMIC_RANGE_NEAR_LIMIT
DICOM Printer
StatusInfo
ELEC DOWN
DICOM Printer
StatusInfo
NORMAL
Diagnostics
Severity
0-Fatal
Severity
3-Notify
4-65