Exposure Transport Belt

Remove the:
Prerequisites:
T10 Torx driver
Tools:
Removal
[1] Remove the:
(a) 5 cover screws
(b) Exposure transport cover
Cover
Screws
[2] Push down on the tensioner to release pressure on the belt.
[3] Slide the belt off the lower shaft and remove the belt.
Installation
Reverse the steps in the removal procedure.
9G9002 – 2011-08-15
Top cover
Right cover
Cover
Screws
Adjustments and Replacements
Belt
Tensioner
Shaft
2-41