SERVICE MANUAL

Film Diverter Assembly

Remove the Felt Pad Assembly
Prerequisites:
Removal
Caution
Hot Surface: When the laser imager is initially de-energized, the processor drum rollers are hot. Take care when
working in the area of the processor.
[1] Remove the 3 screws that hold the film diverter assembly to the cooling plate.
[2] Remove the film diverter assembly.
Film Diverter Assembly
Screws
Installation
[1] Clean any residue in the area before you install the film diverter.
[2] Install the new film diverter assembly with the 3 screws.
Install the felt pad
assembly.
[3]
[4] Close the heat shield and tighten the captive screws.
[5] Install the top cover and the left cover.
[6] Connect the power cord, and then energize the imager.